ArticleJournal of robotic surgery2025
SETY©: A next-generation robotic training platform for safe and autonomous surgery in space - surgical biomechatronics design and Hardware-Software integration within the SY-MIS project.
Article in Journal of robotic surgery, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers, 1 of them a synthesis that pooled it.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed, 1 synthesis or guideline pooled it.
- Multi-centric international 10-year review of surgical magnetic mechatronics in HPB, bariatrics, OB-GYN, urology and neurosurgery: minimally invasive technology, robot design & actuator-sensor fusion.Journal of robotic surgery · 2026Pooled it
- Mechatronic Design and Development of a Lower-Limb Exoskeleton System Based on Knee Joint Biomechanical Principles Using Electro-Pneumatic Actuation with an Embedded EMG Controller for Experimental Validation in Elderly Gait Rehabilitation Support.Bioengineering (Basel, Switzerland) · 2026Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
33 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
introductionSurgical capability in space is constrained by extreme conditions such as microgravity, communication latency, surgical training, and limited resources. Current robotic training platforms are optimized for terrestrial use. The objective is to explore The SETY© platform, developed under the SY-MIS program, designed as a compact and accessible mechatronic surgical platform for minimally invasive procedures with emphasis on operational feasibility in resource-limited and early robotic training exposure to prepare future surgeons for extreme and spaceflight conditions. The final goal is to overcome some barriers for surgeons in their initial phases of training such as earlier exposure to a robotic platform, limited training lab availability, portability issues, on demand practice, affordability of robotic platform, and improving overall surgical proficiency and patient care. DESIGN, SETTING, AND
participantsThis experimental feasibility study (2022-2024) involved the SETY© system in terrestrial analog environments simulating microgravity constraints. The system incorporated ergonomic considerations, low-cost components, and an open-source control architecture to enhance accessibility for academic and clinical training. The platform consisted of dual articulated robotic arms mounted on a compact 400 × 500 mm base, driven by servomechanism motors via a curved rail mechanism. Structural components were produced using 3D-printed PLA and validated via finite element analysis. The electrical system employed an ESP32 microcontroller for console input and an Arduino Nano for arm control, enabling real-time coordination of eight servomotors through potentiometer and joystick inputs. Primary platform outcomes were mechanical stability, kinematic precision (velocity and torque control), synchronization between console and arms, and robustness under repetitive cycles. Secondary platforms outcomes included compactness, energy efficiency, and feasibility for integration into training curricula.
resultsFinite element analysis confirmed a maximum Von Mises stress of 0.482 MPa and minimal deformation (0.0007 mm), yielding a safety factor of 15. The system maintained joint velocities below 10°/s and torques under 10 N·m. Simulation with sinusoidal inputs showed stable motion patterns, velocity peaks at ± 4°/s, torque oscillations around ± 0.05 N·m, and acceleration up to 1500°/s². Electrical integration achieved smooth real-time synchronization without predictive filtering, although the absence of ROS-based frameworks reduced interoperability and limited integration with higher-level control systems. Minor mechanical tolerances and sensor variance were observed but did not compromise functionality.
conclusionsThe SETY© platform demonstrated mechanical robustness, precise motion control, and functional reliability within a compact, energy-efficient design. Its low-cost, open-source configuration supports adoption in surgical and medical engineering education and has potential applicability for space mission surgical preparedness. Future refinements should optimize mechanical damping and material stiffness to enhance performance under high-load or rapid-maneuver conditions.
Indexed as
Identifiers
41272356What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.