Evidence map›Paper›PMID 42853466›Full record

ArticlePhysical and engineering sciences in medicine2026

ACPSEM guideline: commissioning and quality assurance of Magnetic Resonance Imaging Linear Accelerators.

Luis Munoz, Reza Alinaghi-Zadeh, Stephen Gibson, Urszula Jelen, Marcus Powers, Sandra Fisher, Nick Cook, Christopher Colyer, Zoë Moutrie

Abstract read
PubMed Publisher
In one paragraph

Article in Physical and engineering sciences in medicine, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

9 authors.

Luis MunozGenesisCare , Adelaide, SA, Australia. lmunoz@onephysics.com.ORCID http://orcid.org/0000-0003-3605-3457
Reza Alinaghi-ZadehAustin Health, Melbourne, VIC, Australia.ORCID http://orcid.org/0000-0003-0771-2521
Stephen GibsonTownsville University Hospital, Townsville, QLD, Australia.ORCID http://orcid.org/0000-0003-0078-0388
Urszula JelenGenesisCare, Alexandria, Sydney, Australia.ORCID http://orcid.org/0000-0003-3410-7036
Marcus PowersTownsville University Hospital, Townsville, QLD, Australia.ORCID http://orcid.org/0000-0002-4235-6123
Sandra FisherAustin Health, Melbourne, VIC, Australia.
Nick CookChristchurch Hospital, Christchurch, New Zealand.ORCID http://orcid.org/0000-0002-0071-181X
Christopher ColyerGenesisCare , Adelaide, SA, Australia.ORCID http://orcid.org/0009-0003-2788-6163
Zoë MoutrieSouth Western Sydney Cancer Services, New South Wales Health, Sydney, NSW, Australia.ORCID http://orcid.org/0000-0003-3258-5384

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

MRI-guided radiotherapy introduces unique quality assurance (QA) requirements due to magnetic field effects on dose delivery, imaging-dependent workflows, adaptive planning and system specific considerations. Consistent, vendor agnostic guidance is needed to support safe commissioning, clinical implementation, and ongoing performance monitoring across MRI-Linac platforms. This guideline defines minimum recommended MRI-Linac commissioning and QA practices to support consistent and safe clinical operation across services. It was developed by the MRI-Linac QA sub-group of the Magnetic Resonance Imaging Linac Working Group using an evidence-informed, consensus-based approach. Key international standards, task group reports, vendor documentation, and peer-reviewed literature relevant to the MRI-Linac were reviewed. Candidate QA tests were mapped to clinical risks and system functions, then organized into commissioning and routine QA domains with suggested cadence. Recommendations were refined through iterative working group review, with emphasis on implementation and auditability. The resulting commissioning recommendations are grouped into domains addressing key performance and safety areas for MRI-Linacs. Additionally, a minimum test set is proposed with suggested frequencies and tolerances where supported by evidence. Adoption of this guideline supports harmonized MRI-Linac QA practice, improved reproducibility across sites, and enhanced patient safety. The framework is designed to be implementable across platforms and evolve as MRI-guided radiotherapy workflows, vendor capabilities, and the evidence matures.

Indexed as

AcceptanceCommissioningMRIMRI-LinacQuality assurance

Identifiers

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.