Association of Implantation Depth With Hemodynamics and Mechanical Consequences in Small Valve-In-Valve TAVR.
Article in JACC. Asia, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
9 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundValve-in-valve (ViV) transcatheter aortic valve replacement in small surgical bioprostheses is often limited by residual gradients and prosthesis-patient mismatch. Implantation depth is operator-controlled, but depth guidance for different valve combinations remains limited.
objectivesThe authors aimed to evaluate the effects of implantation depth on ViV hemodynamics, leaflet deformation, and pinwheeling in small surgical bioprostheses and to assess the mechanical consequences of THV underexpansion using accelerated wear testing.
methodsA 21-mm self-expanding transcatheter heart valve was implanted in five 21-mm surgical heart valves at multiple depths in a pulse duplicator. Mean gradient, effective orifice area, regurgitation volume, energy loss, leaflet deformation, and pinwheeling were assessed. Selected constrained ViV configurations underwent 200 million cycles of accelerated wear testing under ISO 5840 conditions.
resultsImplantation depth substantially influenced ViV performance. In 3 porcine surgical valves, the most favorable combined hemodynamic profile occurred within a moderate-to-deep depth window of approximately 13.0 to 14.5 mm, although the optimal value differed across individual hemodynamic and deformation endpoints. Excessive depth increased frame constraint, leaflet distortion, and pinwheeling. Two bovine surgical valves showed mixed or U-shaped responses, consistent with differences in true internal diameter and transcatheter valve oversizing. The selected ViV configurations completed 200 million accelerated wear cycles without gross structural failure under bench conditions.
conclusionsIn small ViV-transcatheter aortic valve replacement, the hemodynamic effects of implantation depth are valve-combination specific. A depth-window strategy may improve acute hydrodynamic performance, but clinical translation requires balancing hemodynamic benefit against leaflet deformation and durability-related uncertainty.