Evidence map›Paper›PMID 42783725›Full record

ArticleJournal of functional biomaterials2026

Biomechanical Effects of Horizontal, Vertical, and Combined Misfits in Full-Arch Implant-Supported Titanium Frameworks: A Three-Dimensional Finite Element Analysis.

Hale Arikan Kalayci, Mustafa Baris Guncu

Abstract read
In one paragraph

Article in Journal of functional biomaterials, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

2 authors.

Hale Arikan KalayciDepartment of Prosthodontics, Faculty of Dentistry, Başkent University, Ankara 06790, Turkey.ORCID 0000-0001-8336-2151
Mustafa Baris GuncuDepartment of Prosthodontics, Faculty of Dentistry, Hacettepe University, Ankara 06100, Turkey.

Funding

Başkent University D-DA25/13
6 · The paper itself

Abstract

This study evaluated the effects of misfit type, magnitude, and location on the stress distribution in full-arch screw-retained implant frameworks. A maxillary finite element model with four implants was analyzed under 16 scenarios, characterized by horizontal misfits of 10, 50, 100, and 200 µm; vertical misfits of 10 and 100 µm; and combined 10-10 and 100-100 µm misfits, each positioned anteriorly or posteriorly. Forced seating was simulated using prescribed displacement; no occlusal load or screw preload was applied. Von Mises stresses were evaluated in the framework, occlusal screws, and implants, and principal stresses were assessed in peri-implant bone. Framework stress increased with horizontal misfit magnitude. When the same numerical misfit value (10 or 100 µm) was applied at the same location under otherwise identical model conditions, vertical misfit produced higher framework stress than horizontal misfit, indicating a direction-dependent response associated with different seating deformation modes. The posterior 100-100 µm combined misfit produced the maximum framework (356 MPa), occlusal screw (193 MPa), and implant (250 MPa) stresses. In contrast, the maximum principal stress peaked at the posterior bone site with the anterior 200 µm horizontal misfit (V0 H200 A; 77 MPa), while the most negative minimum principal stress occurred at the posterior bone site with the posterior 200 µm horizontal misfit (V0 H200 P; -52 MPa). Among the tested scenarios, combined misfits yielded the highest framework and screw stresses, whereas implant and bone responses depended on the magnitude and location of the misfit. These findings indicate that the types, magnitudes, and locations of misfits should be considered during framework-fit assessments, with particular attention to combined misfits, before definitive seating.

Indexed as

biomechanical stressfinite element analysisframework misfitimplant-supported prosthesisperi-implant bonetitanium framework

Identifiers

PMID42783725
PMCPMC13608655

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked