ArticleJournal of applied clinical medical physics2026
Clinical evaluation and regression test of a commercial deep-learning auto-segmentation model.
Article in Journal of applied clinical medical physics, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
13 authors.
Funding
Abstract
backgroundCommercial deep-learning segmentation (DLS) tools are increasingly used in clinical practice. Software updates may alter segmentation performance, highlighting the need for systematic clinical evaluation before implementation. PURPOSE: This study presents our experience in clinically evaluating and regression-testing a RayStation DLS model with the aim of guiding commissioning and routine quality assurance of commercial DLS tools in clinical environments.
methodsU-Net DLS model for normal-tissue structures, originally commissioned in RayStation version 11B, was regression-tested after upgrading to 2024A. Previously commissioned CT datasets for head and neck (n = 18), thorax (n = 18), male pelvis (n = 18), and breast (n = 21) were re-evaluated in RayStation 2024A for regression testing following the software upgrade, while 20 new abdominal cases were included for initial commissioning. Clinical segmentations served as the reference standard. Geometric performance was evaluated using Dice Similarity Coefficient (DSC), 95% Hausdorff Distance (HD95), and Mean Surface Distance (MSD). Dosimetric evaluation used the original clinical treatment plans to compare Dmean and Dmax between DLS and clinical contours, normalized to prescription dose; breast was excluded from dosimetric analysis because its DLS model was not commissioned for clinical use.
resultsRegression testing showed that, in comparison with the previously commissioned data, mean [± standard deviation (SD)] DSC values were 0.81 ± 0.09 for the head and neck, 0.84 ± 0.13 for the thorax, and 0.89 ± 0.06 for male pelvis. Corresponding MSD values were 0.83 ± 0.34 mm, 2.47 ± 1.81 mm, and 1.76 ± 0.68 mm; and HD95 values were 4.75 ± 2.38 mm, 8.45 ± 3.38 mm, and 13.21 ± 7.05 mm. Significant paired differences between versions were limited to selected structures, including the mandible, cochlea, and eye in head and neck; lungs and esophagus in thorax; and prostate in the male pelvis (p < 0.05). Breast segmentation results showed low performance (DSC 0.66 ± 0.15, MSD 4.8 ± 1.93 mm, and HD95 22.2 ± 9.6 mm) compared to clinical reference, so the breast structure models are not recommended for clinical use. Abdominal structures (RayStation 2024A vs clinical) had DSC 0.95 ± 0.014; MSD 0.82 ± 0.32 mm; and HD95 5.4 ± 2.9 mm. The mean dose differences between DLS and clinical segmentations were 0.03 Gy for the head and neck, -0.92 Gy for the thorax, 0.27 Gy for the male pelvis, and 0.02 Gy for the abdominal sites, and the corresponding maximum dose differences were 10.96 Gy, 13.23 Gy, 3.39 Gy, and 6.31 Gy.
conclusionsWe described a procedure for practical quality assurance of auto-segmentation tools for clinical use after periodic software upgrades. Regression testing of head and neck, thorax, and male pelvis sites showed consistent performance after the software upgrade and the DLS models for abdominal site were commissioned in this study for clinical use; while the DLS models for breast site were not recommended.