Commissioning and end-to-end validation of a combined surface-guided and triggered kV imaging workflow for breath-hold SBRT on a Varian TrueBeam.
Article in Journal of applied clinical medical physics, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundRespiratory motion introduces significant geometric uncertainty in Stereotactic Body Radiotherapy (SBRT) for thoracic and abdominal tumors. Deep Inspiration Breath-Hold (DIBH) mitigates this, but the surface signal alone may not reflect internal target position, supporting the need for real-time internal-anatomy verification during delivery. PURPOSE: To commission a Surface-Guided Radiation Therapy (SGRT) system and validate an integrated SGRT + Image-Guided Radiation Therapy (IGRT) + triggered kV imaging (SITI) workflow for DIBH SBRT. MATERIALS &
methodsThe LAP LUNA 3D SGRT system was commissioned per AAPM TG-302 using phantom-based assessment of static and dynamic localization accuracy, reproducibility, and latency. An end-to-end test using a dynamic phantom validated the SITI workflow. Point dose was measured with an ion chamber under four scenarios: SGRT-only delivery, full SITI delivery, and SITI with induced 1 and 2 mm uncorrected 3D shifts.
resultsStatic localization accuracy was better than 0.5 mm / 0.3°, with reproducibility within 0.2 mm / 0.1°. Dynamic testing confirmed sub-millimeter spatial accuracy and a 31.6 ms latency, well below the AAPM TG-302 100 ms tolerance. The 1 and 2 mm uncorrected shifts produced point-dose reductions of 0.9% and 2.2% relative to the SITI reference, consistent with the ∼1%/mm local dose gradient. Both shifts were clearly visualized on triggered kV images, confirming detection of sub-tolerance residual displacements undetected by surface guidance alone.
conclusionsThe LAP LUNA 3D system meets the technical requirements for SBRT. The SITI workflow is technically feasible and provides real-time visualization of internal target position during delivery, offering a robust motion-management strategy for DIBH SBRT of mobile thoracic and abdominal targets.