ArticleDiving and hyperbaric medicine2026
Evaluation of external ventricular drain performance under hyperbaric conditions: a bench study.
Article in Diving and hyperbaric medicine, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
Introduction: External ventricular drains (EVDs) are widely used for intracranial pressure (ICP) monitoring and cerebrospinal fluid (CSF) diversion in neurocritical care. Hyperbaric oxygen treatment (HBOT) exposes medical devices to increased ambient pressure; however, limited data exists regarding EVD performance and reliability under hyperbaric conditions. Methods: A bench study was conducted to evaluate EVD performance under hyperbaric conditions. A rigid, non-compressible saline-filled column was used to assess pressure measurement stability independent of compliance effects. Additional compliance-based syringe models were explored to assess the feasibility of evaluating unclamped drainage behaviour. A commercially available EVD system (CODMAN EDS 3; Codman and Shurtleff, a Johnson and Johnson company, Raynham, MA, USA) was exposed to stepwise ambient pressures from 101 to 283 kPa (1.0-2.8 atmospheres absolute) in a multiplace hyperbaric chamber. Pressure measurements, drainage behaviour, fluid column stability, and system integrity were assessed. Results: The CODMAN EDS 3 EVD system maintained structural integrity across tested pressure levels. While pressure measurements were stable in the rigid open glass syringe model, the closed compliance-based models demonstrated pressure variability during chamber pressurisation, including when closed to fluid drainage. Unclamped testing using these models demonstrated inconsistent fluid dynamics, precluding reliable assessment of fluid drainage behaviour under hyperbaric conditions. Conclusions: The CODMAN EDS 3 EVD system remained structurally intact during hyperbaric exposure; however, pressure measurements varied in closed models, including when closed to fluid drainage. ICP measurements obtained during HBOT should be interpreted with caution due to potential pressure-related artefacts. Reliable continuous assessment of fluid pressure drainage behaviour remains limited by current bench model constraints. Further research incorporating advanced brain compliance simulation models and carefully designed in-vivo observational studies is required to better characterise EVD behaviour under hyperbaric conditions and to ensure safe clinical use and accurate interpretation of ICP measurements in the hyperbaric environment.