Transitioning Beyond Mercury Sphygmomanometers: Diagnostic Accuracy and Agreement of Digital Blood Pressure Monitors in a South Indian Adult Population.

Binu Areekal, Anju C Mathew, Manu Naish

Abstract read
In one paragraph

Article in Cureus, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors.

Binu AreekalDepartment of Community Medicine, Government Medical College Ernakulam, Kochi, IND.
Anju C MathewDepartment of Community Medicine, Government Medical College Ernakulam, Kochi, IND.
Manu NaishDepartment of Pulmonology, Government Medical College Ernakulam, Kochi, IND.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

introductionHypertension is a leading cause of morbidity and mortality, and its prevalence is increasing worldwide. Accurate blood pressure (BP) measurement is fundamental to clinical diagnosis and epidemiological surveillance, and any measurement errors can lead to substantial misclassification of hypertension, resulting in inappropriate clinical management and inaccurate estimates of disease burden. While automated digital monitors offer a practical alternative to manual devices, their real-world accuracy requires robust local validation during the phase-out of traditional instruments. The objective of the current study was to assess the validity and accuracy of a digital BP monitor against the gold-standard mercury device in an adult population in South India. MATERIALS AND

methodsA cross-sectional diagnostic evaluation study was conducted among 262 adults attending the outpatient department of a teaching hospital in South India. Baseline BP was measured twice using the digital BP monitor (Omron HEM-7120, OMRON Healthcare Co., Ltd., Kyoto, Japan) with a one- to two-minute interval between readings. After a five-minute interval, BP was measured using a mercury sphygmomanometer (Novaphon, NOVAFON GmbH, Weinstadt, Germany). Validity, reliability, agreement, and bias were analyzed using paired t-tests, Pearson correlation, and Bland-Altman analysis.

resultsDigital and mercury measurements demonstrated exceptionally strong linear correlations for systolic BP (r = 0.945) and diastolic BP (r = 0.845). The mean difference (digital minus mercury) was non-significant for both parameters: -0.167 ± 7.46 mmHg for systolic and 0.296 ± 6.63 mmHg for diastolic. When comparing the two devices, absolute differences within 5 mmHg were observed in 135 (51.5%) participants for systolic BP and 164 (62.6%) participants for diastolic BP. Categorical assessment showed substantial diagnostic agreement (Cohen's kappa = 0.77), with a sensitivity of 84.0% and specificity of 92.9% for hypertension detection. Bland-Altman analysis revealed a negligible mean bias of -0.167 mmHg for systolic BP (95% limits of agreement (LoA): -14.79 to 14.46 mmHg) and 0.296 mmHg for diastolic BP (95% LoA: -12.70 to 13.29 mmHg).

conclusionsThe automated digital monitor demonstrates good population-level agreement and diagnostic validity, making it a reliable tool for routine screening and epidemiological surveillance. However, given the wide limits of agreement, critical clinical decisions, especially in borderline cases, should be made with caution.

Indexed as

agreementbland–altman analysisblood pressure monitoringdiagnostic accuracydigital bp monitorhypertensionmercury sphygmomanometersouth indiavalidity