Evidence map›Paper›PMID 42724569›Full record

ArticleJournal of thoracic disease2026

Comparison of Y20 with GEN11CN (Harmonic HD 1000i) ultrasonic energy device for sealing 5-7 mm pulmonary vessels in thoracoscopic surgery: a multicenter, randomized, assessor- and statistician-blinded, noninferiority trial.

Weijia Wu, Zhangqiang Chen, Yumei Shen, Qifeng Ding, Xuejun Xu, Xing Jin, Yongsheng Gong, Ning Zhong, Shanzhou Duan, Yongbing Chen

Abstract read
In one paragraph

Article in Journal of thoracic disease, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

10 authors.

Weijia Wu *Department of Thoracic Surgery, The Second Affiliated Hospital of Soochow University, Suzhou, China.
Zhangqiang Chen *Department of Thoracic Surgery, The Second Affiliated Hospital of Soochow University, Suzhou, China.
Yumei Shen *Department of Thoracic Surgery, The Second Affiliated Hospital of Soochow University, Suzhou, China.
Qifeng DingDepartment of Thoracic Surgery, The Second Affiliated Hospital of Soochow University, Suzhou, China.
Xuejun XuDepartment of Thoracic Surgery, The Second Affiliated Hospital of Soochow University, Suzhou, China.
Xing JinDepartment of Thoracic Surgery, The Second Affiliated Hospital of Soochow University, Suzhou, China.
Yongsheng GongDepartment of Thoracic Surgery, Suzhou Municipal Hospital, Suzhou, China.
Ning ZhongDepartment of Thoracic Surgery, The First People's Hospital of Kunshan, Suzhou, China.
Shanzhou DuanDepartment of Thoracic Surgery, The Second Affiliated Hospital of Soochow University, Suzhou, China.
Yongbing ChenDepartment of Thoracic Surgery, The Second Affiliated Hospital of Soochow University, Suzhou, China.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Background: Reliable sealing of 5-7 mm pulmonary vessels remains a persistent challenge in thoracoscopic surgery. Endostaplers are difficult to maneuver within the narrow interlobar fissure, and vascular clips are prone to slippage that can obscure the operative field. Ultrasonic energy devices offer an alternative for vessels of this caliber, but direct clinical comparisons between different device platforms are lacking. This study aimed to compare the success rate of sealing 5-7 mm blood vessels between a novel device (Y20) and the widely used GEN11CN (Harmonic HD 1000i). Methods: This was a multicenter, randomized, assessor- and statistician-blinded non-inferiority study. We enrolled patients aged 18-80 years scheduled for thoracoscopic anatomic segmentectomy or lobectomy who had at least one 5-7 mm vessel requiring sealing on computed tomography (CT) images, and randomly assigned them 1:1 to Y20 group or GEN11CN group. Patients with coagulopathy, severe thrombocytopenia, or recent active bleeding were excluded. Randomisation was stratified by centre using a central, computer-generated sequence held by an independent statistician. Device branding was masked with opaque covers so that the outcome evaluator remained blinded. The primary endpoint was the intraoperative vessel sealing success rate, defined as the absence of stump bleeding within a 10-minute observation window, with a prespecified non-inferiority margin of -10%. Secondary endpoints included intraoperative blood loss, operative time, postoperative drainage volume, and complications. Safety outcomes included laboratory parameters, vital signs, device-related sealing failures, and device-attributable complications. Results: Baseline characteristics were comparable except for higher systolic blood pressure (BP; 126.16 Conclusions: Y20 was noninferior to GEN11CN for intraoperative sealing of 5-7 mm pulmonary vessels and showed a comparable safety profile, supporting Y20 as a clinically acceptable alternative for this specific indication. These findings should not be generalized to vessels >7 mm or to settings without immediate rescue capability. Trial Registration: Chinese Clinical Trial Registry (ChiCTR2400088694; https://www.chictr.org.cn/showproj.html?proj=223068).

Indexed as

hemostasisPulmonary vesselsrandomized controlled trialthoracic surgery, video-assistedultrasonic surgical procedures

Identifiers

PMID42724569
PMCPMC13559264

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.