Monte Carlo simulation and experimental validation of SPECT systems via energy spectrum decomposition and image profile analysis.
Article in EJNMMI physics, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundAccurate modelling and performance evaluation of gamma-ray scintillation detectors are required in single photon emission computed tomography (SPECT) systems to ensure reliable nuclear medicine diagnostics. Monte Carlo simulation is therefore a powerful computational technique used to accurately model the gamma-ray scintillation detectors and photon transport in medical imaging systems. Therefore, the objective of this study is to validate the SIMIND Monte Carlo simulation code using a self-developed peak-detection algorithm for clinical SPECT gamma cameras (Siemens E.Cam and GE Infinia), with a focus on energy-spectrum decomposition and image-profile analysis.
methodsThis study presents a dual-system evaluation of clinical SPECT gamma cameras-the Siemens E.Cam and GE Infinia systems using the SIMIND Monte Carlo simulation code. We assessed collimator and detector-specific SPECT performance metrics, including energy resolution, sensitivity, image evaluation, spatial resolutions, and spectral fidelity for a
resultsEnergy resolution differences ≤ 4% and sensitivity variations < 1.2% were observed between simulated and experimental measurements. Spatial resolution analysis revealed a range of 3.5-4.8 mm for Siemens E.Cam (bar and pie phantoms) and ~ 1 cm for GE Infinia, highlighting system-specific imaging capabilities. A custom algorithm successfully identified key spectral components, including photopeaks, Compton edges, and lead X-ray peaks, with deviations < 3%. Image profile analysis revealed FWHM values of 4-4.5 pixels for both systems.
conclusionThis work shows that SIMIND accurately reproduces the fundamental photon interaction characteristics including energy spectra, energy resolution, and sensitivity for both Siemens E.Cam and GE Infinia SPECT systems. However, for image-profile evaluation, good agreement was obtained for Siemens E.Cam, whereas the GE Infinia results revealed a system-specific discrepancy associated with the actual clinical condition of the installed camera. Nevertheless, when used appropriately, SIMIND provides a robust framework for optimizing quantitative SPECT protocols and guiding system design improvements, particularly for well-calibrated systems.