ArticleGraefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie2026
Repeatability and reproducibility of corneal endothelial measurements using the topcon SP-1P specular microscope.
Article in Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeTo evaluate the repeatability and reproducibility of corneal endothelial parameters measured with the Topcon SP-1P specular microscope in healthy eyes, comparing single and panoramic acquisition modes.
methodsThis reproducibility study included 137 eyes from 79 healthy adults. Endothelial cell density (ECD), average cell area, coefficient of variation (CV), hexagonality, and pachymetry were measured twice by two independent examiners using both single and panoramic modes. Statistical agreement was assessed using intraclass correlation coefficients (ICC), repeatability coefficients (RC), and Bland-Altman analysis.
resultsIntra-observer reproducibility for ECD, average cell area, and pachymetry was excellent in both modes (ICC ≥ 0.95). However, the panoramic mode consistently showed improved precision, with higher ICCs and lower RCs for most morphometric variables. Inter-observer agreement was also excellent for ECD (ICC 0.94) and average cell area (ICC 0.95). Significant systematic differences were found between acquisition modalities; specifically, ECD was on average 7.27% lower in single mode compared to panoramic mode (p < 0.001).
conclusionThe Topcon SP-1P demonstrates high reproducibility for primary endothelial parameters, particularly when using the panoramic acquisition mode. Since systematic differences exist between modalities, single and panoramic measurements are not directly interchangeable, highlighting the importance of standardizing techniques for longitudinal clinical monitoring.