ArticleMedical physics2026
Megavoltage photon minibeam production through the embedded multi-leaf collimator of a clinical linac: A dosimetric characterization.
Article in Medical physics, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
Abstract
backgroundAlthough spatially-fractionated radiation therapy (SFRT) has been receiving increasing clinical interest, its complete and optimized clinical transfer requires several technological and radiobiological issues to be addressed. PURPOSE: The aim of this work is to setup and dosimetrically characterize an experimental platform generating planar megavoltage photon minibeams to exploit quantitative radiobiological SFRT experiments.
methodsMinibeam patterns were generated on a Varian TrueBeam™ linear accelerator equipped with a high definition (HD) MLC, which consists of 60 tungsten leaf pairs, whom the 32 inner ones project a thickness of 2.5 mm at the isocenter plane (i.e., 100 cm distance from the radiation source) and the 28 outers of 5 mm. Each pattern was obtained alternating opening one leaf after closing one, two, or three of the smallest MLC leaves. Measurements were performed with a PTW microdiamond detector in edge-on orientation in a water phantom over depths from 1 to 30 cm, varying the beam energy and the source to surface distance (SSD) in different sessions. A Python-based in-house automated fitting procedure was used to obtain from the measured profiles all the minibeam parameters: peak widths (PWs), distance between consecutive peaks (CTCs), peak to valley dose ratios (PVDRs) and depth doses.
resultsThe best performances were obtained at SSD = 70 cm and 6 MV, where PWs and PVDRs respectively range from 1.2 to 2.4 mm and from 2 to 7. For all patterns the average dose varies in depth analogously to the reference field, PVDRs show an almost steady decrease except at shallower depths, while PWs and CTCs widen linearly with the depths. Moreover, their variation over different sessions was within the estimated uncertainty of the single measurement.
conclusionsWe are claiming that a Varian TrueBeam™ equipped with an HD-MLC represents a practical, stable, reproducible and accessible platform that might be suitable for radiobiological studies on SFRT and for the investigation of the relative underlying mechanisms.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.