ArticleNeuroradiology2026
Clinical outcome after rescue intracranial stenting: a comparison of the ACCLINO flex and the CREDO heal stenting platforms after failed mechanical thrombectomy.
Article in Neuroradiology, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
17 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeMechanical thrombectomy (MT) is the first-line therapy for patients with acute ischemic stroke due to large vessel occlusion; however, despite substantial advances in MT techniques and devices, mechanical recanalization remains unsuccessful in up to 15% of patients. Intracranial rescue stenting represents a promising option to achieve successful recanalization and good clinical outcomes in patients with failed MT.
methodsIn this retrospective, observational single-center analysis 77 consecutive patients who underwent failed MT and a maximum of 4 unsuccessful device passes were included. The ACCLINO flex stent / CREDO
resultsA clinically favorable outcome was achieved in 64.1% (25/39) of patients treated with the CREDO heal stent and in 39.5% (15/38) of patients treated with the ACCLINO flex stents (p = 0.031). Rescue stenting with the CREDO heal stent (aOR 3.51 95% CI 1.11-11.13, p = 0.033), baseline NIHSS (aOR 0.97 95% CI 0.85-0.97, p = 0.005) and pmRS (aOR 0.52 95% CI 0.31-0.89, p = 0.016) were independently associated with a favorable clinical outcome. Baseline characteristics did not differ between both groups.
conclusionsThis study suggests that the design of the implanted stent device may influence the clinical outcome. Features such as optimized radial force and thrombogenicity-reducing surfaces of stents used for RIS may improve clinical outcomes of patients treated with intracranial rescue stenting in unsuccessful MT. These findings are hypothesis-generating and require confirmation in prospective, randomized studies.
Indexed as
Identifiers
42684435What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.