Development and performance evaluation of an upright dedicated cone-beam breast CT system.
Article in Medical physics, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Development and performance evaluation of an upright dedicated cone-beam breast CT system.Medical physics · 2026Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
9 authors.
Funding
Abstract
backgroundDedicated breast CT is an emerging breast X-ray imaging modality. While current commercial breast CT systems use prone-patient, pendant-breast geometry, the system described here uses upright patient geometry with the uncompressed breast supported by a cup. PURPOSE: The purpose of this work is to describe the development of a newly designed, upright geometry, dedicated cone-beam breast CT system and to evaluate its imaging performance using objective metrics.
methodsThe prototype system uses a tungsten-target, mammography-format, X-ray tube operating at 60 kV with 0.25 mm Cu and 1 mm Al added filtration, and a complementary metal-oxide semiconductor (CMOS) detector with 0.152 mm pixel pitch coupled to 500 microns thick CsI:Tl scintillator. During short scan acquisition, the X-ray source moves inferior to the breast, and 210 projections are acquired over an angular range of 210 degrees. The projections are reconstructed to an isotropic voxel pitch of 0.22 mm using Feldkamp-Davis-Kress (FDK) algorithm with Parker weights. Quantitative performance measures including linearity, modulation transfer function (MTF), and noise power spectrum (NPS) were evaluated. Phantom studies were conducted at various X-ray tube current (mA) and pulse-width (ms) combinations with the objective of determining the minimum detectable size of low-contrast targets and calcium carbonate spheres representing soft tissue lesions and microcalcification clusters, respectively.
resultsThe measured 1st HVL was 4.23 ± 0.01 mm of Al. The limiting resolution (10% MTF) was 2.18 mm
conclusionsThe developed upright breast CT system showed the ability to achieve high spatial resolution and low contrast resolution. The excellent technical performance of the breast CT system reported here suggests that further investigations using patient imaging are warranted.