Dosimetric impact of aperture-restricted VMAT on heart dose in breast cancer treated with regional nodal irradiation.
Article in Journal of applied clinical medical physics, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
17 authors.
Funding
Abstract
backgroundVolumetric modulated arc therapy (VMAT) is increasingly used for breast and chest wall (CW) treatment with regional nodal irradiation (RNI) because of its ability to sculpt dose around complex target volumes. However, VMAT can increase low-dose radiation to surrounding normal tissues and particularly increase cardiac dose. PURPOSE: To evaluate the dosimetric impact of a strategic aperture-restricted beam geometry, compared with traditional open-field geometry in VMAT planning for comprehensive left-sided breast/CW irradiation with RNI across three linac platforms. The impact of convergence mode settings was also investigated.
methodsFourteen patients with left-sided breast/CW irradiation including comprehensive RNI were retrospectively included. For each patient, a patient-specific optimization template with a prescription of 40.05 Gy in 15 fractions was applied across three linac platforms (Varian TrueBeam, Varian Halcyon, and Elekta Versa HD), using two beam geometries: an open-field approach, in which field sizes fully encompassed the targets, and an aperture-restricted approach, in which the deep field border was limited to reduce beam entry to the heart and ipsilateral lung. Jaw tracking was enabled on the TrueBeam and Versa HD linacs. Each plan was also evaluated across three convergence modes ("Off," "On," and "Extended"), resulting in 18 plans per patient. All plans were normalized such that 95% of the breast/CW planning target volume received 95% of the prescription dose. Dosimetric endpoints include mean heart dose, ipsilateral lung V17Gy, contralateral lung V4Gy, and contralateral breast/CW V5Gy. Total monitor units (MUs) were evaluated, and patient-specific quality assurance (PSQA) was performed for the three highest-MU plans on each linac platform. Paired tests were used to compare across plans.
resultsTarget coverage was maintained after plan normalization. Aperture-restricted VMAT significantly reduced mean heart dose compared with open-field geometry across all linac platforms and convergence modes. The proportion of plans exceeding mean heart dose of 4 Gy was substantially lower with aperture-restricted geometry compared to open-field plans (2.4% vs 26.2%). The largest reduction in mean heart dose was observed with the "Off" convergence mode, decreasing from 4.20 to 2.71 Gy on TrueBeam, 3.32 to 2.58 Gy on Halcyon, and 3.97 to 2.84 Gy on Versa HD. Ipsilateral lung V17Gy was also significantly reduced with aperture-restricted beam geometry, with only one exception on Halcyon using the "Extended" mode. No significant differences were observed in contralateral lung V4Gy, although contralateral breast/CW V5Gy increased significantly, representing a dosimetric tradeoff associated with the aperture-restricted geometry. Aperture-restricted beam geometry significantly increased MUs; however, PSQA passed for the three highest-MU plans on each linac platform according to our clinical criteria. Additionally, more extensive convergence modes were associated with improved plan quality, particularly in reducing hotspot dose (D0.1cc) and enhancing cardiac and ipsilateral lung sparing.
conclusionsAperture-restricted beam geometry significantly reduced mean heart dose and ipsilateral lung dose while maintaining target coverage across all three linac platforms, supporting its potential clinical value in VMAT planning for comprehensive left-sided breast/CW irradiation with RNI, albeit with increased contralateral breast/CW dose. When combined with a more extensive convergence mode, overall plan quality was further improved.