OPT-former: A transformer-based approach for eye motion tracking in ocular proton therapy.
Article in Medical physics, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
11 authors.
Funding
Abstract
backgroundOcular proton therapy (OPT) is the main conservative treatment option for ocular melanoma but requires precise gaze stability monitoring during irradiation. In current clinical workflows at the Centro Nazionale di Adroterapia Oncologica (CNAO) in Pavia, this task is performed manually by trained operators, which can introduce reaction delays and human error. PURPOSE: To propose and assess the performance of OPT-former, a transformer-based method for real-time segmentation of the iris and pupil, designed for integration into automated safety interlock systems.
methodsThe model, inspired by the MedFormer architecture, performs direct three-class segmentation (background, iris, pupil) on 256 × 256 images acquired by CNAO's Eye Tracking System. A dataset of 1683 frames from 173 patients was split into independent training (1349 frames), validation (110 frames), and testing (224 frames) sets. Data augmentation was applied to increase variability, resulting in 3902 training images. Performance was evaluated using Intersection over Union (IoU), Dice coefficient, and Szymkiewicz-Simpson coefficient (SSC), assessing the OPT-former performance and comparing them with a previously proposed 3-step U-Net and manual expert segmentations. Patient-specific fine-tuning was implemented using pre-treatment dry-run session videos.
resultsThe OPT-former achieved segmentation accuracy comparable to the 3-step U-Net, but with substantially lower inference times (8.62 Hz on GPU vs. 3.26 Hz), enabling near frame-by-frame real-time analysis. Patient-specific fine-tuning significantly improved all metrics for both structures (p < 0.05), with the largest gains for the iris. After fine-tuning, the model outperformed manual annotations from two experts under simulated clinical conditions.
conclusionsThe proposed OPT-former achieves high segmentation accuracy, with reduced pipeline complexity and inference time suitable for clinical application. These results endorse its role as a robust support tool in clinical practice, seamlessly integrating into established workflows without impacting the consolidated routine, and enabling automated gaze monitoring and real-time safety interlock activation in OPT.