ArticleActa orthopaedica2026
Wear performance of metal and ceramic long-neck femoral head taper connections: an experimental ex vivo study.
Article in Acta orthopaedica, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
background and purposeThe wear performance of femoral head taper connections is important for the longevity of total hip replacement. Metal release may cause adverse tissue reactions. There are some clinical indications that long neck heads may show increased wear, but no laboratory tests have been published. We aimed to examine the wear performance of femoral head taper connections in long and extra-long neck heads.
methodsLoad frame and hip joint simulator wear tests were performed with the same stem designs as in the reference study. The CoCr and zirconia toughened alumina (ZTA) heads of 36 mm diameter were of the long or extra-long neck type. Optical and scanning electron microscopy were utilized to study wear marks, and mass spectrometry to estimate Ti and Co release. Head disassembly forces were measured.
resultsOur preclinical, design-specific results showed that, compared with medium necks, long and especially extra-long necks led to increased wear while maintaining an asymmetric wear pattern more widely distributed throughout the contact area. In the hip joint simulator tests, 1 neck fracture occurred.
conclusionThe long and extra-long head taper connections showed increased wear compared with that of medium heads. Although no serious damage occurred, CoCr head tapers were the most affected compared with ZTA.
Indexed as
Identifiers
What OpenQuestion holds