Clinical implementation and performance evaluation of a motorized MR-compatible OCTAVIUS 4D system for SBRT patient-specific QA on a 1.5 T MR-LINAC.

Clement Chevillard, Marie Blanchet, Cyrielle Mella, Robin Sauvinet, Rezart Belshi, Ivan Taranenko

Abstract readEvaluation Study
In one paragraph

Article in Journal of applied clinical medical physics, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Clement ChevillardDepartment of Radiation Oncology, Service of Medical Physics, Institut Curie, Saint Cloud, France.ORCID https://orcid.org/0009-0004-5698-8502
Marie BlanchetDepartment of Radiation Oncology, Service of Medical Physics, Institut Curie, Saint Cloud, France.
Cyrielle MellaDepartment of Radiation Oncology, Service of Medical Physics, Institut Curie, Saint Cloud, France.
Robin SauvinetDepartment of Radiation Oncology, Service of Medical Physics, Institut Curie, Saint Cloud, France.
Rezart BelshiDepartment of Radiation Oncology, Service of Medical Physics, Institut Curie, Saint Cloud, France.
Ivan TaranenkoElekta Limited, Cornerstone, Crawley, West Sussex, UK.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundMagnetic resonance-guided radiotherapy using 1.5 T MR-LINAC systems introduces specific constraints for patient-specific quality assurance (PSQA), requiring MR-compatible dosimetric devices capable of accurate three-dimensional dose verification. While previous studies have characterized the dosimetric performance of MR-compatible detector arrays, the clinical implementation of rotational three-dimensional QA systems in MR-guided workflows remains insufficiently documented. PURPOSE: This study investigated the clinical implementation and practical performance of a motorized MR-conditional OCTAVIUS 4D system for stereotactic body radiotherapy (SBRT) PSQA on a 1.5 T MR-LINAC.

methodsThe MR-conditional OCTAVIUS 4D phantom equipped with a 1500 MR ionization chamber array was implemented for PSQA measurements on a Unity 1.5 T MR-LINAC (Elekta, Stockholm, Sweden). A total of 40 clinical step-and-shoot treatment plans were analyzed with an SBRT prescription. The system was evaluated in terms of three-dimensional dose reconstruction accuracy under clinically relevant off-axis condition, angular positioning consistency with the gantry positioning, and integration into clinical workflow. Measured dose distributions were compared with treatment planning system (TPS) calculations using a 3D gamma analysis with a local 2.5%/2.5 mm criterion and a 10% dose threshold. Independent angular measurements were additionally performed using a digital inclinometer.

resultsThe system demonstrated high agreement between measured and calculated dose distributions, with a mean gamma passing rate of 98.3% (local 2.5%/2.5 mm). All plans met institutional acceptance criteria (GPR ≥ 95%). Independent angular measurements demonstrated excellent agreement between gantry angle, phantom angle, and software-reported angle, with deviations below 0.6° and no systematic bias observed. No visible dependence of gamma passing rate on radial off-axis target positioning was observed within the investigated clinical range. The evaluated beam arrangements included irregular angular spacing and clockwise/counterclockwise gantry transitions during delivery, supporting effective synchronization between the OCTAVIUS 4D MR system and the Unity gantry under clinical step-and-shoot conditions.

conclusionsThe motorized MR-conditional OCTAVIUS 4D system demonstrated clinically feasible implementation for rotational three-dimensional PSQA for SBRT on a 1.5 T MR-LINAC. The system provided consistent dose reconstruction accuracy and reliable angular positioning under clinically relevant delivery conditions, supporting its integration into routine MR-guided radiotherapy QA workflows.

Indexed as

Magnetic Resonance ImagingParticle AcceleratorsPhantoms, ImagingQuality Assurance, Health CareRadiosurgeryRadiotherapy, Image-GuidedRadiotherapy Planning, Computer-AssistedHumansRadiotherapy DosageRadiotherapy, Intensity-ModulatedMR‐LINACOCTAVIUS 4D phantomOCTAVIUS detector 1500 MRpatient‐specific quality assurance

Identifiers

PMID42573267
PMCPMC13455677

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.