ArticleCase reports in anesthesiology2026

Underrecognized Interaction Between PICO Dressings and Pacemakers: A Case From the ICU.

Reginald Caldecott, John Laffey

Abstract read
In one paragraph

Article in Case reports in anesthesiology, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

2 authors.

Reginald CaldecottDepartment of Anaesthesia and Intensive Care Medicine, University Hospital Galway, Galway, Ireland, hse.ie.ORCID https://orcid.org/0000-0001-8296-2999
John LaffeyDepartment of Anaesthesia, School of Medicine, University of Galway, Galway, Ireland, universityofgalway.ie.ORCID https://orcid.org/0000-0002-1246-9573

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Background: Modern pacemakers incorporate a magnet-responsive switch that, on exposure to a sufficient static magnetic field, converts the device to fixed-rate asynchronous (magnet-mode) pacing. This is designed behaviour rather than malfunction. Interactions between this feature and magnet-containing medical equipment are underrecognised. The PICO 7 negative pressure wound therapy (NPWT) system contains a magnet in its pump unit, which the manufacturer states can interfere with nearby pacemakers. Case Presentation: An 81-year-old man was admitted to the High Dependency Unit after right hemicolectomy. He had atrial fibrillation with a slow ventricular response and a permanent pacemaker programmed VVI at 60 beats per minute. Overnight, telemetry showed a sustained paced rate of 100 beats per minute; he was asymptomatic. The PICO 7 pump unit was found resting on his chest, over the pacemaker generator. Repositioning the pump to the foot of the bed returned the paced rate to 60 beats per minute immediately, and device interrogation the following day confirmed normal pacemaker function. Discussion: The fixed rate of 100 beats per minute, its dependence on pump position and immediate resolution on removal are characteristic of magnet-mode activation rather than generic electromagnetic interference. Although the patient was asymptomatic, a sustained rate of 100 beats per minute carried a risk of myocardial ischaemia and heart failure decompensation given his cardiac history. Recognition and a simple bedside intervention resolved the episode. Conclusion: Magnet-containing medical equipment placed near a cardiac implantable electronic device can trigger magnet-mode pacing. The risk depends on the position of the magnet-bearing component, not on the wound site. An unexplained change in a paced rate should prompt early consideration of a nearby magnetic source, because the remedy may be immediate and at the bedside.

Indexed as

asynchronous pacingcardiac implantable electronic deviceICU safetymagnet modenegative pressure wound therapypacemakerPICO dressing

Identifiers

PMID42558917
PMCPMC13438167

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.