Evidence map›Paper›PMID 42550258›Full record

ArticleEuropean journal of trauma and emergency surgery : official publication of the European Trauma Society2026

Comparison of fluoroscopic versus electromagnetic distal locking in humeral shaft osteosynthesis using intramedullary nail: a retrospective study.

Roman Madeja, Jiri Szeliga, Jiri Voves, Jana Pometlova, Adéla Konde, Jaroslav Janosek, Pavel Dousa

Abstract readComparative Study
PubMed Publisher
In one paragraph

Article in European journal of trauma and emergency surgery : official publication of the European Trauma Society, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Roman MadejaDepartment of Trauma Surgery, University Hospital Ostrava, 17. listopadu 1790, Ostrava, 708 52, Czech Republic. roman.madeja@fno.cz.ORCID http://orcid.org/0000-0002-9253-3116
Jiri SzeligaDepartment of Trauma Surgery, University Hospital Ostrava, 17. listopadu 1790, Ostrava, 708 52, Czech Republic.
Jiri VovesDepartment of Trauma Surgery, University Hospital Ostrava, 17. listopadu 1790, Ostrava, 708 52, Czech Republic.ORCID http://orcid.org/0000-0002-8352-8198
Jana PometlovaDepartment of Trauma Surgery, University Hospital Ostrava, 17. listopadu 1790, Ostrava, 708 52, Czech Republic.ORCID http://orcid.org/0000-0002-0733-915X
Adéla KondeDepartment of Deputy Director for Science and Research, University Hospital Ostrava, 17. listopadu 1790, Ostrava, 708 52, Czech Republic.ORCID http://orcid.org/0000-0002-0511-0749
Jaroslav JanosekCentre for Health Research, Faculty of Medicine, University of Ostrava, Syllabova 19, Ostrava, 703 00, Czech Republic.ORCID http://orcid.org/0000-0003-0682-4521
Pavel DousaDepartment of Trauma Surgery, University Hospital Ostrava, 17. listopadu 1790, Ostrava, 708 52, Czech Republic.ORCID http://orcid.org/0000-0002-2499-421X

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeIntramedullary nailing is a standard method for osteosynthesis of diaphyseal humeral fractures. Distal locking, however, typically requires fluoroscopic guidance, which increases radiation exposure. This study aimed to compare standard fluoroscopy-guided distal locking with the Ezy-Aim system (Austofix, Australia) based on electromagnetic pulse detection from the perspectives of surgical times, radiation exposure, and complications.

methodsA retrospective study included 58 patients treated for diaphyseal humeral fractures (AO/OTA 12A, 12B) between 2019 and 2023. Group A (n = 28) received osteosynthesis using the Ezy-Aim system; Group B (n = 30) underwent fluoroscopy-guided distal locking. Data on operative time, fluoroscopy time, radiation dose, and complications were analysed.

resultsIn all, distal locking was not successful at the first attempt in 14.3% (4/28) of screws placed with Ezy-Aim and 20.0% (6/30) when fluoroscopy was used. Time and dose analyses were performed on the remaining 48 patients. The total surgical time was significantly longer with the Ezy-Aim system (median 58 min) compared to fluoroscopy (median 52 min, p = 0.034), primarily due to the time required for the device assembly. However, distal screw placement was faster with Ezy-Aim (median 12 min vs. 14 min, p = 0.019). Radiation exposure was significantly lower in the Ezy-Aim group (median total dose 418 cGy/cm

conclusionsThe Ezy-Aim system reduces radiation exposure and speeds up distal screw placement, though it prolongs total operative time due to assembly requirements. This method offers a safe and effective alternative to fluoroscopic guidance for distal locking in humeral shaft fractures.