ArticleJournal of orthopaedic translation2026

Optimising bionic pore designs in a spiral blade implant to enhance bone ingrowth and stability: A combined CFD and large-animal study.

Meng-Xuan Yao, Tai-Long Shi, Yu-Qin Zhang, Yi-Fan Zhang, Chuan Ren, Shi-Wu Dong, Zhi-Yong Hou, Ying-Ze Zhang, Hong-Zhi Lv, Wei Chen

Abstract read
In one paragraph

Article in Journal of orthopaedic translation, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

10 authors.

Meng-Xuan YaoDepartment of Orthopaedic Surgery, Hebei Medical University Third Hospital, Shijiazhuang, Hebei, 050051, China.
Tai-Long ShiDepartment of Orthopaedic Surgery, Hebei Medical University Third Hospital, Shijiazhuang, Hebei, 050051, China.
Yu-Qin ZhangDepartment of Orthopaedic Surgery, Hebei Medical University Third Hospital, Shijiazhuang, Hebei, 050051, China.
Yi-Fan ZhangDepartment of Orthopaedic Surgery, Hebei Medical University Third Hospital, Shijiazhuang, Hebei, 050051, China.
Chuan RenDepartment of Orthopaedic Surgery, Hebei Medical University Third Hospital, Shijiazhuang, Hebei, 050051, China.
Shi-Wu DongDepartment of Biomedical Materials Science, College of Biomedical Engineering, Army Medical University (Third Military Medical University), Chongqing, 400038, China.
Zhi-Yong HouDepartment of Orthopaedic Surgery, Hebei Medical University Third Hospital, Shijiazhuang, Hebei, 050051, China.
Ying-Ze ZhangDepartment of Orthopaedic Surgery, Hebei Medical University Third Hospital, Shijiazhuang, Hebei, 050051, China.
Hong-Zhi LvDepartment of Orthopaedic Surgery, Hebei Medical University Third Hospital, Shijiazhuang, Hebei, 050051, China.
Wei ChenDepartment of Orthopaedic Surgery, Hebei Medical University Third Hospital, Shijiazhuang, Hebei, 050051, China.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Purpose: This study aimed to evaluate the effects of different Bionic Pore geometries (circular, triangular, and rectangular) and diameters (0.8 mm, 1 mm, 2 mm, and 3 mm) on fluid shear stress (FSS) distribution and bone ingrowth in a bionic spiral blade implant. Methods: Twelve Bionic Pore structures were modeled using CAD software, and computational fluid dynamics (CFD) analysis was performed to simulate the FSS distribution in a bone environment. Based on these simulation results, the circular hole design was selected and manufactured via electron beam melting (EBM). It was then incorporated into a bionic spiral blade (length: 50 mm, outer diameter: 6 mm, and a central canal of 2 mm for surgical pin placement). Four female sheep (weighing 35-40 kg) were implanted with these spiral blades, divided into 3-month and 6-month follow-up groups. After surgery, X-ray imaging, micro-CT, and histological evaluation were used to assess implant stability, bone bridging, and tissue response within the bionic pores. Results: CFD simulations showed that circular pores generated a more optimal and uniform FSS distribution; the 0.8 mm circular pore produced an average FSS of 0.0086 ± 0.00013 Pa, representing a 79.1% and 68.0% reduction compared to identically sized triangular (0.0412 ± 0.0176 Pa) and rectangular pores (0.0269 ± 0.0056 Pa), respectively. In vivo experiments demonstrated that small-diameter holes significantly accelerated osteogenesis. At 3 months, the 0.8 mm group achieved a bone tissue area ratio (BTA/BHA) of 15.44 ± 1.94%, which was 98.7% higher than that of the 3 mm group (7.77 ± 3.03%). By 6 months, the 0.8 mm group reached a BTA/BHA of 39.79 ± 7.80%, demonstrating a 131.2% greater bone occupancy compared to the 3 mm group (17.21 ± 3.97%, Conclusion: This study indicates that optimizing the geometry and size of bionic pores can improve FSS conditions and promote bone ingrowth. Small-diameter circular holes, in particular, favor early and robust osseointegration. Moreover, this work underscores the potential of combining advanced manufacturing techniques with mechanobiological principles to design customized bionic implants for enhanced orthopedic outcomes. Importantly, the close agreement between CFD predictions and in vivo outcomes validates computational fluid dynamics as a practical prescreening tool for optimizing pore designs in load-bearing implants. The translational potential of this article: This "bench-to-bedside" research has established a robust framework for clinical translation. Leveraging the aforementioned optimized parameters, a novel bionic internal fixation device has been developed, which has successfully secured a U.S. utility patent and obtained a medical device registration certificate. This bionic device facilitates trabecular bone ingrowth and remodeling in parallel with cortical bone healing. Consequently, the findings of this study provide a validated, evidence-based foundation for the future design and development of bionic orthopedic implants.

Indexed as

Bionic poresBone ingrowthFluid shear stressInternal fixationLarge animal modelOsseointegration