In-house development of CT-based 3D-printed head & neck phantom for Radiotherapy applications.
Article in Journal of applied clinical medical physics, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundAnthropomorphic phantoms are widely used in Radiotherapy for quality assurance (QA), treatment verification and dosimetric validation purposes. However, commercially available anthropomorphic phantoms are expensive and offers limited customization, restricting their broader clinical use. The advancement three-dimensional (3D) printing technology has provided a versatile and cost-effective approach for fabrication of such patient-specific anthropomorphic phantoms with accurate anatomical and radiological properties. PURPOSE: This study aims to design, fabricate and evaluate a CT-based Head & Neck anthropomorphic phantom using 3D printing technique for Radiotherapy applications.
methodsThe design of the phantom is developed from the computed tomography (CT) images of a male patient. Relevant organs such as brain, eyes, trachea, thyroid, skull, mandible, spine, soft tissue etc were segmented and reconstructed. Suitable tissue-equivalent materials compatible with 3D printing were selected and optimized to replicate the radiological properties of the human tissues. CT scan images of the fabricated phantom were acquired using a CT simulator. Hounsfield Unit (HU) values of each organ was compared with the patient CT data to assess tissue equivalence and uniformity. For dosimetric evaluation of the phantom, the acquired CT images of the fabricated phantom were imported into the treatment planning system (TPS). A treatment plan was generated in TPS to deliver 2 Gy dose to planning target volume (PTV) and executed in a 6MV Linear accelerator. Dose verification was performed by placing thermoluminescent dosimeters (TLDs) at predefined locations within the PTV and organs at risk (OARs). Measured doses were compared with TPS-calculated dose values to evaluate dosimetric accuracy.
resultsThe developed phantom demonstrated good agreement between the patient CT data and the phantom CT images. The agreement of HU values across analyzed slices was within 5%, indicating accurate replication of tissue-equivalent attenuation properties. Intra-organ HU uniformity was also within 5%, confirming consistent material composition and printing quality. Dosimetric comparison revealed that deviations between TPS-calculated and TLD-measured doses were less than 5% within the PTV, 1%-5% for OARs, and 7%-9% at distal exit dose measurement points. The PTV received approximately 190cGy, showing good agreement between planned and delivered doses. Larger percentage deviations observed at distal exit points, where doses were below 10 cGy, were attributed primarily to detector uncertainties and increased measurement uncertainty at low dose levels rather than inaccuracies in phantom design or treatment delivery.
conclusionThe developed 3D-printed CT-based indigenous Head & Neck phantom accurately replicates both the anatomical and tissue equivalence characteristics of the patient CT data. The phantom demonstrated reliable dosimetric parameters supporting its suitability for Radiotherapy applications. The phantom is customizable, simple to use, cost effective and could be a promising alternative to commercial anthropomorphic phantoms.