A preliminary comparison of prosthetic socket liner strain determined using digital image correlation and finite element analysis.

Mohammadreza Freidouny, Michael L Madigan, Carson Squibb, Masaki Hada, Abbie Bailey, Trevor Johnson, Brian Kaluf, Michael K Philen

Abstract readComparative Study
In one paragraph

Article in PloS one, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors.

Mohammadreza FreidounyGrado Department of Industrial & Systems Engineering, Virginia Tech, Blacksburg, Virginia, United States of America.ORCID https://orcid.org/0009-0002-4827-7239
Michael L MadiganGrado Department of Industrial & Systems Engineering, Virginia Tech, Blacksburg, Virginia, United States of America.ORCID https://orcid.org/0000-0002-4299-3851
Carson SquibbKevin T. Crofton Department of Aerospace and Ocean Engineering, Virginia Tech, Blacksburg, Virginia, United States of America.
Masaki HadaKevin T. Crofton Department of Aerospace and Ocean Engineering, Virginia Tech, Blacksburg, Virginia, United States of America.
Abbie BaileyKevin T. Crofton Department of Aerospace and Ocean Engineering, Virginia Tech, Blacksburg, Virginia, United States of America.
Trevor JohnsonVirginia Prosthetics and Orthotics, Christiansburg, Virginia, United States of America.
Brian KalufSalt Lake City Research Hub, Ottobock, Salt Lake City, Utah, United States of America.
Michael K PhilenKevin T. Crofton Department of Aerospace and Ocean Engineering, Virginia Tech, Blacksburg, Virginia, United States of America.ORCID https://orcid.org/0000-0002-3526-8811

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Improving prosthetic socket fit remains an important issue among prosthesis users. Understanding strains at limb-socket interface may help prosthetists improve socket fit. Previous studies have used finite element analysis (FEA) to predict stresses and strains at limb-socket interface. However, the selection of FEA model material properties as well as the validation of FEA strain results remain challenging. Digital image correlation (DIC) is a well-accepted optical technique that is widely used in experimental mechanics to noninvasively measure strains and can be used to validate FEA predictions. The goal of this feasibility study was to demonstrate the use of DIC to measure prosthesis liner strains of a unilateral transtibial prosthesis user and conduct a preliminary comparison of these strains with those estimated from an FEA model. One participant with a transtibial residual limb was recruited. Liner strains of the residual limb during two stepping tasks were measured with DIC. An FEA model of the residual limb was also developed, and its estimates of liner strains were compared with DIC measurements. Results showed that mean and probability distributions of principal strain values measured by DIC were in moderate to good agreement with FEA estimations. For example, DIC measured mean maximum and minimum principal strains of 0.041 and 0.021 in the anterior aspect of the residual limb when it was being vertically loaded, and FEA estimated mean maximum and minimum principal strains of 0.051 and 0.027 in the same region. The Bhattacharyya coefficient (a measure of overlap between two probability distributions, with values approaching 1 indicating high similarity) exceeded 0.90 for principal strain distributions between DIC and FEA on the anterior aspect of the residual limb during both tasks. These findings demonstrate that DIC can be used to refine FEA model material properties and results. As such, this methodology may help to advance the design of prosthetic sockets; however, as this study involved a single participant, findings are preliminary and intended to demonstrate methodological feasibility rather than generalize across the broader population of prosthesis users.

Indexed as

Artificial LimbsFinite Element AnalysisBiomechanical PhenomenaHumansProsthesis DesignStress, Mechanical

Identifiers

PMID42447121
PMCPMC13367698

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.