Evidence map›Paper›PMID 42397419›Full record

ArticleStrahlentherapie und Onkologie : Organ der Deutschen Rontgengesellschaft ... [et al]2026

Effects of intrafractional motion on targeting accuracy in frameless single-isocenter multi-target stereotactic radiosurgery.

Maximilian Grohmann, Cordula Petersen, Katharina Hintelmann, Manuel Todorovic

Abstract read
PubMed Publisher
In one paragraph

Article in Strahlentherapie und Onkologie : Organ der Deutschen Rontgengesellschaft ... [et al], 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors.

Maximilian GrohmannDepartment of Radiotherapy and Radiation Oncology, University Medical Center Hamburg-Eppendorf, Hamburg, Germany. m.grohmann@uke.de.ORCID http://orcid.org/0000-0002-6909-811X
Cordula PetersenDepartment of Radiotherapy and Radiation Oncology, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
Katharina HintelmannDepartment of Radiotherapy and Radiation Oncology, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
Manuel TodorovicDepartment of Radiotherapy and Radiation Oncology, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeThis study evaluates the impact of intrafractional motion on targeting accuracy in frameless single-isocenter multi-target stereotactic radiosurgery (SIMT) for brain metastases and examines the necessity of intrafractional six-degrees-of-freedom (6DoF) image-guided corrections.

methodsA total of 120 SIMT-SRS treatments and 294 couch kicks were analyzed using positioning data from an intrafractional stereoscopic X‑ray image-guidance system. Deviations exceeding clinical tolerance (0.5 mm/0.5°) triggered repositioning. Initial and final 6DoF shifts were compared. The dosimetric impact of simulated setup errors (0.5-1 mm, 0.5°-1°) was evaluated in 15 selected SIMT-SRS cases comprising 90 metastases. Target coverage and plan quality were assessed using D98%, conformity index, and gradient index. A multivariable regression model was applied to assess the influence of target volume and distance to the isocenter on target coverage.

resultsIntrafractional motion exceeded tolerance thresholds after 66% of couch kicks, necessitating corrections. Translational and rotational errors decreased from 0.71 ± 0.38 mm and 0.58 ± 0.34° before correction to 0.33 ± 0.19 mm and 0.38 ± 0.18° after correction. Simulated setup errors progressively impaired target coverage and conformity. For 0.5 mm/0.5° errors, D98% decreased by 7.6% ± 2.5%, conformity index declined by 9.5% ± 4.9%, and gradient index increased by 1.2% ± 2.3%. At 1 mm/1°, D98% decreased by 17.3% ± 4.8%, conformity index declined by 26.6% ± 9.2%, and gradient index increased by 1.1% ± 2.6%. Smaller targets (≤ 0.4 cm

conclusionIntrafractional motion in frameless SIMT-SRS frequently exceeds clinical tolerance thresholds after couch kicks and can compromise both target coverage and conformity. Even small setup errors degrade plan quality, particularly for small lesions, whereas changes in dose gradient are comparatively modest within the investigated range. Intrafractional 6DoF corrections are therefore essential to reduce geometric deviations and preserve treatment quality.