ArticleSpine deformity2026
An in vitro study to identify worst-case parameters for assessing mechanical performance of anterior vertebral body tethering (AVBT) systems.
Article in Spine deformity, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
Abstract
purposeAnterior vertebral body tethering (AVBT) is a novel surgical option for scoliosis patients that preserves spinal mobility. However, tether-related failures such as tether breakage are common, and cause of these failures is unknown. Furthermore, there is a lack of data on how different design and surgical factors influence the subsequent in vivo performance of these AVBT systems. This study aims to examine the mechanical durability of a mock AVBT system under various mechanical loading scenarios resulting from important surgical factors.
methodsA mock AVBT system was designed using Titanium screws and Polyethylene Terephthalate (PET) tether to evaluate mechanical performance under various clinically relevant loading conditions. Static and dynamic tensile tests assessed the effects of screw angulation (0°/40°), tether gauge length (length between the vertebral anchors: 25/76 mm), setscrew torque (5/6/7 Nm), and retightening. Constructs were tested using Instron equipment; measuring stiffness, yield, and failure loads. Statistical analyses compared configurations to identify factors influencing AVBT durability.
resultsFor this mock AVBT system, failure primarily occurred at the tether/setscrew junction due to slippage. Greater screw angulation (40°) increased yield and runout loads compared to 0°. Changing the gauge length had minimal effect on failure loads. Furthermore, the higher setscrew torque (7 Nm) improved strength and fatigue resistance, while setscrew retightening also increased yield and runout loads.
conclusionsDue to the infancy of this technology, there is a need for deeper understanding of the failure mechanics of AVBT systems. These findings identify testing parameters that affect mechanical performance and suggest future research directions for assessing the mechanical integrity of AVBT devices.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.