ArticleSurgical endoscopy2026
Changes in ultrasonic scalpel performance and characterization following gastric cancer surgery.
Article in Surgical endoscopy, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundThis study aims to investigate the performance degradation and morphological changes of ultrasonic scalpels following gastric cancer surgery, and to evaluate their potential interactions with medical-grade titanium alloys (such as titanium clips and staplers).
methodsThree experimental groups were established: Group C (control group), Group OP (post-gastric cancer surgery group), and Group OP+Ti (post-gastric cancer surgery group exposed to titanium alloys). Cutting rates were measured for each group and analyzed using one-sided t-tests and linear mixed-effects models. Surface morphology was observed using scanning electron microscopy (SEM).
resultsCompared with Group C, all blades in Group OP exhibited varying degrees of tissue carbonization, adhesion, and microcracks, with a significant decrease in cutting rate (P < 0.05). Group OP+Ti not only displayed the aforementioned phenomena observed in Group OP but also exhibited severe morphological damage at the blade tip, with a similarly significant reduction in cutting rate. Interaction with titanium alloys accelerated blade wear, further compromising their performance.
conclusionsDuring gastric cancer surgery, the cutting performance and characteristics of ultrasonic scalpels undergo significant degradation. Their interaction with titanium alloy implants is a critical factor that cannot be overlooked, as it may cause additional damage to the blade tip, leading to a decline or even loss of performance. It is recommended that surgeons use the instruments properly, avoid contact with metal during ultrasonic scalpel operation, and regularly assess the wear status and safety of the instruments to ensure patient safety during surgery and optimal postoperative recovery.