Influence of various disinfection methods on the flexural strength of additively manufactured denture base resins: an in vitro study.
Article in BMC oral health, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
10 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundWith the advancement of technology for digitally fabricated removable dental prostheses, protocols for the disinfection of three-dimensional (3D)-printed dentures are lacking. Therefore, this study aimed to assess and compare the effects of different disinfection protocols on the flexural strength (FS) of 3D-printed denture base resins. MATERIALS AND
methodsA total of 200 bar-shaped samples (64 × 10 × 3.3 mm) were fabricated via two types of 3D-printed denture base resins: DentaBase (DB) and Denture 3D+ (D). According to the disinfectant method, each resin sample was divided into five groups (n = 20). The samples were then further subdivided based on the duration of disinfection, either 90 days or 180 days. One control group was kept in water, while the other four groups were disinfected using a effervescent denture cleanser tablets (EDC), microwave (MW), ultraviolet light (UV), or ultrasonic (US). The flexural strength (FS) was measured. The collected data were statistically analyzed via three-way ANOVA, and post hoc Tukey's test was used for data analysis (α = 0.05).
resultsFor Denture 3D+, FT significantly reduced FS at 90- and 180-day intervals (p < 0.05). UV disinfection maintained the highest FS at 90 days (72.1 MPa, p < 0.013); however, both the US and MW methods resulted in significant reductions by 180 days (p < 0.05). In the DentaBase, FT and MW disinfection caused significant FS reductions at 90 days (p < 0.001). By 180 days, FT, US, and MW disinfection methods led to significant decreases in FS (p < 0.001). However, UV light did not have a significant effect on FS at either point (p = 0.22).
conclusionUV disinfection proved to be the most suitable method for maintaining the mechanical integrity of 3D-printed denture base resins, while chemical, microwave, and ultrasonic methods should be used with caution due to their deteriorating effects on flexural strength.