Evidence map›Paper›PMID 42260482›Full record

ArticleBMC oral health2026

Automated micro-CT quantification of clear aligner fit: a pilot comparison of manufacturing processes.

Shuyan Liu, Ming Lv, Junyu Li, Miguizhi Zhou, Yukang Zhang

Abstract readComparative Study
In one paragraph

Article in BMC oral health, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

Shuyan Liu *Department of Stomatology, Center for Plastic & Reconstructive Surgery, Zhejiang Provincial People's Hospital (Affiliated People's Hospital), Hangzhou Medical College, Hangzhou, Zhejiang, 310014, China.
Ming Lv *The First People's Hospital of Lin'an District, Hangzhou, Zhejiang, 311300, China.
Junyu LiGuilin University of Electronic Technology, Guilin, Guangxi, 541004, China.
Miguizhi ZhouXianju Traditional Chinese Medicine Hospital, Taizhou, Zhejiang, 317300, China.
Yukang ZhangXianju Traditional Chinese Medicine Hospital, Taizhou, Zhejiang, 317300, China. m13013319965@gmail.com.

Funding

This work was supported by the Medical Science and Technology Project of Zhejiang Province under Grant 2025KY663
6 · The paper itself

Abstract

backgroundClear aligner fit accuracy significantly influences treatment efficacy. Micro-computed tomography (micro-CT) enables precise three-dimensional quantification of aligner-tooth adaptation, but manual measurement is time-intensive and operator-dependent.

objectiveTo develop and validate an automated flood-fill algorithm for micro-CT gap quantification and compare fit accuracy across three aligner manufacturing processes: in-house 3D-printing, Invisalign (thermoformed), and Angelalign (thermoformed).

methodsThis in vitro pilot study used a standardized maxillary dental model to fabricate nine aligners across three manufacturing groups (three aligners per group): direct DLP 3D-printing (LuxCreo Inc., Chicago, US), Invisalign (SmartTrack™), and Angelalign (Angel Pro™). Each aligner was scanned using a Xradia 610 Versa micro-CT (ZEISS, Oberkochen, Germany) at 58.82 μm voxel resolution. An automated flood-fill algorithm was developed in Python to segment and quantify aligner-tooth gap widths from reconstructed volumes, and validated against manual segmentation using Dice similarity coefficient, sensitivity, and specificity. Group differences in gap width were assessed using the Kruskal-Wallis test with Mann-Whitney post-hoc comparisons and Bonferroni correction.

resultsThe automated algorithm achieved a mean Dice similarity coefficient of 0.87 ± 0.04 (range: 0.81-0.94), sensitivity of 0.89 ± 0.05, and specificity of 0.96 ± 0.03. A total of 120 gap width measurements were obtained (3D-printing: n = 42, Invisalign: n = 42, Angelalign: n = 36). Mean gap widths were 193.27 ± 93.03 μm, 186.96 ± 81.86 μm, and 254.89 ± 133.21 μm, respectively. The Kruskal-Wallis test showed no significant overall difference among groups (H = 4.396, P = 0.111). Post-hoc analysis revealed no significant difference between 3D-printing and Invisalign (U = 886.0, P = 0.974, Cliff's Delta = 0.005, negligible effect). Neither comparison with Angelalign reached statistical significance after Bonferroni correction (P = 0.070 and P = 0.065), though both showed small effect sizes (Cliff's Delta ≈ - 0.23). Linear mixed-effects models confirmed these findings (random-effects ICC = 0.179 for aligner-level clustering).

conclusionsThe automated flood-fill algorithm may enable efficient and reproducible micro-CT gap quantification, reducing processing time from ~ 45-60 min to approximately 3.2 min per scan. No statistically significant difference in fit accuracy was observed between in-house 3D-printed and Invisalign thermoformed aligners (mean difference: 6.31 μm), providing preliminary evidence for their potential as a cost-effective manufacturing alternative; however, this should not be interpreted as evidence of equivalence given the small aligner sample size. The Angelalign group showed a trend toward larger gaps with small effect sizes, warranting further investigation. Substantial within-group variability across all manufacturing methods underscores the need for systematic quality control in aligner production.

Indexed as

Orthodontic Appliance DesignOrthodontic Appliances, RemovableX-Ray MicrotomographyAlgorithmsHumansImaging, Three-DimensionalPilot ProjectsPrinting, Three-Dimensional3D-printingClear alignersFit accuracyFlood-fill algorithmImage segmentationMicro-CT

Identifiers

PMID42260482
PMCPMC13573369

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.