Performance evaluation of the binary MLC system on a novel biology-guided radiotherapy machine using beamlet sequence log data from daily clinical treatments.
Article in Journal of applied clinical medical physics, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
Abstract
purposeTo evaluate the clinical performance of a binary multi-leaf collimator (MLC) that makes high-frequency transitions on a novel ring-gantry medical linac.
methodsBeamlet sequence data from treatment plans and daily treatments were retrieved for ten patients treated with image-guided radiotherapy (IGRT) on the novel machine, with five patients treated in the pelvic and thoracic regions respectively. The prescription dose ranged from 1.8 to 7 Gy per fraction with an average beam-on time of 484.1 ± 129.6 s (range: 299-666 s). All plans used a field size of 40 cm x 2 cm. A 64-leaf binary MLC modulates the beam at 50 gantry firing positions in each gantry rotation while the gantry rotates at 60 RPM. Gantry angle, couch position, delivered monitor units (MU), and MLC leaf opening status were recorded at each beam firing position. Daily beamlet sequence log data was compared with those in the treatment plan at each couch position.
resultsThe ten treatment plans had an average MU per fraction of 5269 ± 1462 MU (range: 3184-7259) with an average total gantry firing positions of 15 835 ± 4391 (range: 9566-21 817). In daily delivery, an average number of 29.9 ± 8.9 (range: 12.3-41.0) gantry firing positions were skipped, resulting in an average reduction of 9.9 ± 3.0 MU (range: 4.1-13.6 MU), or 0.19 ± 0.04% (range: 0.13-0.28%), relative to plan MUs. The MU reduction was proportional to the number of skipped gantry firing positions, while the number of skipped gantry firing positions had an approximate linear relationship with the total number of gantry firing positions in a plan.
conclusionIn-house software tools have been developed to efficiently evaluate daily performance of the binary MLC on the novel machine. In delivering IGRT plans, the number of skipped gantry positions or MU reduction showed no clinically significant effect on delivered dose.