Hemocompatibility of Mechanical Circulatory Support Devices: An Integrated Review of Platelet Function Assessment in In-Vitro Blood Experiments.

Xiangyu He, Michael B Fischer, Daniel Zimpfer, Alice Assinger, Marcus Granegger

Abstract readReview
PubMed Publisher
In one paragraph

Review in Annals of biomedical engineering, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

Xiangyu HeChristian Doppler Laboratory for Mechanical Circulatory Support, Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, AKH Wien, Währinger Gürtel 18-20, A-1090, Wien, Vienna, Austria.
Michael B FischerClinic for Transfusion Medicine and Cellular Therapy, Medical University of Vienna, Vienna, Austria.
Daniel ZimpferChristian Doppler Laboratory for Mechanical Circulatory Support, Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, AKH Wien, Währinger Gürtel 18-20, A-1090, Wien, Vienna, Austria.
Alice AssingerLudwig Boltzmann Institute of Cardiovascular Research, Vienna, Austria.
Marcus GraneggerChristian Doppler Laboratory for Mechanical Circulatory Support, Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, AKH Wien, Währinger Gürtel 18-20, A-1090, Wien, Vienna, Austria. marcus.granegger@meduniwien.ac.at.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeDespite persistent bleeding and thrombotic adverse events in patients supported with blood pumps, standardized preclinical assessment of hemocompatibility remains largely restricted to hemolysis testing. To advance the field, robust and comparable in vitro platelet function testing within mock-loop mechanical circulatory support (MCS) studies is needed. This review systematically summarizes current methodologies for platelet function assessment in MCS devices and critically examines their advantages and limitations to inform the development of a unified, evidence-based testing framework.

methodsA systematic literature search was conducted in PubMed, Web of Science, Embase, and Scopus using Medical Subject Headings (MeSH) and free-text terms in accordance with PRISMA guidelines. Articles were screened according to predefined inclusion and exclusion criteria.

resultsThirty-nine studies met the eligibility criteria. Study designs varied substantially with respect to experiment duration, sampling strategies, mock-circulation loop configurations, and pump types. Human whole blood was the predominant test medium (n = 27). Anticoagulation strategies differed widely, with citrate-based formulations (n = 22) and heparin (n = 15) most frequently employed. Flow cytometry was the predominant analytical method (n = 25), primarily assessing platelet function through changes in surface receptor expression or shedding.

conclusionFlow cytometry offers a highly informative and direct approach for assessing platelet state in MCS research compared to other methods. However, substantial methodological heterogeneity continues to hinder data interpretation and cross-study comparability. Beyond standardizing experimental setups and protocols, future investigations should incorporate multiple complementary biomarkers and clarify their relationships to clinically relevant outcomes to more comprehensively characterize shear-induced platelet responses.

Indexed as

Blood experimentin vitro experimentsMechanical circulatory support devicePlatelet functionShear stress