Evidence map›Paper›PMID 42233100›Full record

ArticleJournal of shoulder and elbow arthroplasty2026

Early radiographic findings are not associated with instability for reverse total shoulder arthroplasty with the Perform metaphyseal-fixated humeral stem.

Amir Boubekri, Sai Devana, Ling Chen, Melissa A Wright

Abstract read
In one paragraph

Article in Journal of shoulder and elbow arthroplasty, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed.

  1. Article
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors.

Amir BoubekriDepartment of Orthopedic Surgery, Washington University School of Medicine, St. Louis, MO, USA.
Sai DevanaDepartment of Orthopedic Surgery, Washington University School of Medicine, St. Louis, MO, USA.
Ling ChenInstitute for Informatics, Data Science, and Biostatistics, Washington University School of Medicine, St. Louis, MO, USA.
Melissa A WrightDepartment of Orthopedic Surgery, Washington University School of Medicine, St. Louis, MO, USA.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Background: Humeral stems in reverse total shoulder arthroplasty (rTSA) have moved toward shorter metaphyseal fixed stems in recent years to preserve bone stock and reduce the risks of stress shielding and stem loosening. However, shorter stems may be at higher risk for stem malalignment, which may impact clinical outcomes such as instability. The purpose of this study was to examine early radiographic outcomes and their relationship to instability, utilizing the Perform metaphyseal-fixated humeral stem for rTSA. Methods: This is a retrospective review of a consecutive series of patients undergoing primary rTSA by a single surgeon for the 2-year period after adoption of a new shorter metaphyseal fixed humeral stem. Standardized radiographs were analyzed at 2 weeks post-operatively and at final follow-up (minimum 1 year). Initial radiographs were analyzed for humeral stem alignment, metaphyseal fill ratio, and distance from stem tip to greater tuberosity. Final x-rays were analyzed for stem alignment, distance from stem tip to greater tuberosity, and humeral loosening. Subsidence of the humeral implant was calculated based on vertical and angular stem changes between the initial and final radiographs. Logistic regression was used to analyze surgical and radiographic parameters associated with subsidence and instability. Results: One hundred seventy-two patients were included in the analysis. The mean neck shaft angle (NSA) for initial post-operative x-rays was 136.4 (standard deviation, 4.5) and for final x-rays was 135.4° (standard deviation, 4.2). The metaphyseal fill ratio was significantly associated with NSA on both initial and final x-rays, with an increased ratio associated with higher NSA in both instances ( Conclusion: In this study of rTSA with the Perform metaphyseal-fixated humeral stem, radiographic parameters including stem alignment and subsidence were not predictors of instability. No stems had radiographic loosening. Established patient risk factors for instability, including diagnosis of rotator cuff tear, remain key risk factors even with newer metaphyseal fixed stems. These findings highlight the importance of patient selection, understanding the nuances of the system, and refining surgical technique when using a new implant to optimize outcomes.

Indexed as

Humeral neck shaft angleHumeral stem malalignmentInstabilityMetaphyseal humeral stemReverse total shoulder arthroplastyShort humeral stemSubsidence

Identifiers

PMID42233100
PMCPMC13223865

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.