Reducing Motion Artifact in High Resolution 7 T MRI Using the Magnetic Resonance Minimal Motion ("MR-MinMo") Head Stabilization Device.
Article in Magnetic resonance in medicine, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed.
- Revisiting the auditory hemodynamic response function in the era of fast fMRI.bioRxiv : the preprint server for biology · 2026Article
- AutoNeuro: An Open-Source fMRI Toolbox for Real-Time Neuroadaptive Task Design.bioRxiv : the preprint server for biology · 2026Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
11 authors.
Funding
Abstract
purposeTo evaluate the effectiveness of the MR-MinMo head stabilization device in mitigating motion artifacts during long-duration, high-resolution 7 T MRI scans, with and without retrospective motion correction
methodsThe MR-MinMo was tested on seven pediatric and 12 adult healthy volunteers using 0.6 mm isotropic, 3D Multi-Echo Gradient Echo (ME-GRE) scans. A ∼10-min ME-GRE scan (linear sampling acceleration factor 2 × 2) and a ∼20-min scan (DISORDER sampling acceleration factor 1.4 × 1.4) were obtained. Both scans were acquired with and without the MR-MinMo in a 2 × 2 factorial study design. Qualitative and quantitative assessment of image quality of the first echo image was obtained via visual inspection and the normalized gradient squared (NGS) metric, respectively. A repeated measures ANOVA was used to determine individual and combined effects of the device and DISORDER images with retrospective motion correction on NGS values, with age as a binary factor (pediatric/adult). T2* maps were generated and the standard deviation of white matter R2*(=1/T2*) values assessed across conditions.
resultsThe MR-MinMo significantly reduced motion artifacts both visually and in terms of NGS scores, particularly in pediatric volunteers. There was a significant interaction between MR-MinMo and DISORDER motion correction, suggesting that MR-MinMo improved retrospective motion correction. T2* maps demonstrated improved visual appearance and reduced WM variance with the MR-MinMo.
conclusionThe MR-MinMo can improve image quality via motion reduction in high-resolution 7 T scans. By keeping motion within a correctable regime, the device can also improve the performance of retrospective motion correction methods.