ArticleEJNMMI physics2026
Quantitative accuracy of
Article in EJNMMI physics, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
9 authors.
Funding
Abstract
backgroundRing-shaped systems based on CZT are promising for quantitative imaging, but their accuracy may be affected by septal penetration of high-energy photons-given the fixed collimator-and by limitations in scatter correction due to low-energy tailing. This study compares the quantitative performance of a ring-shaped CZT system with that of a conventional dual-head NaI SPECT/CT system (StarGuide™ and NM/CT 870 DR, both GE HealthCare) using the 208 keV photopeak of
methodsA NEMA phantom with 3D-printed anthropomorphic inserts, including kidneys and lumbar vertebrae, as well as standard spherical inserts, was filled with activity concentrations mimicking the distribution observed in patients 1 day post-injection of 7.4 GBq of [
resultsThe total activity measured within the phantom deviated from the expected value by means of 13% for the StarGuide™ and 0% for the NM/CT 870 DR. After partial-volume-effect correction based on recovery coefficients, the activity concentration in kidney inserts was measured with ~ 5% accuracy on both systems. For the StarGuide™, mean errors in tumour inserts (volumes: 2.6-28 ml) ranged from - 8% to 3%, whereas a consistent underestimation by an average of - 11% was observed for the NM/CT 870 DR. The StarGuide™ significantly overestimated the background activity concentration (59%), while the NM/CT 870 DR showed a smaller error (- 8%). Reproducibility was lower for the StarGuide™, with maximum CV values of 7%, compared with 3% for the NM/CT 870 DR. Clinical findings were consistent with phantom results.
conclusionsThe StarGuide™ provided slightly more accurate activity quantification in regions of high relative uptake, such as tumours and kidneys, compared with the NM/CT 870 DR. However, quantification in regions of low relative uptake and total activity estimation were less accurate, likely due to septal penetration. The lower reproducibility suggests that longer scan durations may be considered.