Impact of Different Pre-Dilution Methods on Routine Blood Test Results: A Study With Sysmex XN-1000.

Lingsong Wu, Mengzhe Qin, Zhe Xing, Fengli Jiang, Sien Zhan, Jintian Hu

Abstract read
In one paragraph

Article in Journal of clinical laboratory analysis, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Lingsong WuDepartment of Medical Laboratory, Plastic Surgery Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
Mengzhe QinDepartment of Cosmetic Injection Center, Plastic Surgery Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
Zhe XingDepartment of Medical Laboratory, Plastic Surgery Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
Fengli JiangDepartment of Medical Laboratory, Plastic Surgery Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
Sien ZhanDepartment of Medical Laboratory, Plastic Surgery Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
Jintian HuDepartment of Cosmetic Injection Center, Plastic Surgery Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.ORCID https://orcid.org/0000-0003-2036-2343

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

objectivesTo compare three manual pre-dilution procedures with the automated undiluted mode on the Sysmex XN-1000 analyzer and to identify an optimal standard operating procedure for pre-dilution testing.

methodsA total of 60 residual EDTA-K3 whole-blood specimens were consecutively collected between March and May 2024 and analyzed within 2 h. Each specimen was tested in the automated undiluted mode and by three manual pre-dilution procedures: direct mixing, dry-tip wiping, and wet-tip wiping. Pre-dilution was performed at 1:7 (20 μL of blood plus 120 μL of Sysmex DCL diluent) using calibrated pipettes. All procedures were performed by one trained technician. Each method was tested in triplicate, and mean values were analyzed. White blood cell count (WBC), red blood cell count (RBC), hemoglobin (HGB), platelet count (PLT), and hematocrit (HCT) were evaluated by paired t-test, Pearson correlation, Passing-Bablok regression, and bias analysis against CLIA allowable total error (TEa) criteria.

resultsFor WBC, RBC, HGB, and PLT, no significant difference was found between any pre-dilution method and the undiluted mode (all p > 0.05); for HCT, only Method 1 differed significantly (p = 0.022). Passing-Bablok regression showed that the wet-tip wiping method had the best overall agreement with the undiluted mode. TEa-based analysis showed that only Methods 2 and 3 met allowable bias limits for all five parameters.

conclusionThe wet-tip wiping method showed the best overall analytical performance and may be preferred for low-volume specimens on the Sysmex XN-1000 platform.

Indexed as

Hematologic TestsBlood Cell CountHematocritHemoglobinsHumansPlatelet CountReproducibility of ResultsHemoglobinsBland–Altman analysisblood routine testpre‐dilution methodsSysmex XN‐1000wet‐tip wiping

Identifiers

PMID42226695
PMCPMC13399759

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.