Peripheral instability gradient in macular thickness measurements across five optical coherence tomography systems: a prospective cross-sectional study of inter-device agreement and clinical monitoring risk.
Article in BMC medical imaging, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
2 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundCross-platform variability in optical coherence tomography (OCT) macular thickness measurements poses a clinically meaningful challenge for longitudinal patient assessment, particularly when patients are imaged on different devices over time. This study characterizes intra-device repeatability and inter-device agreement across five contemporary OCT systems, and introduces the Peripheral Instability Gradient (PIG)-the progressive deterioration in cross-device measurement concordance from the foveal center outward across ETDRS-defined macular zones.
methodsIn this prospective, cross-sectional study, 38 healthy adults underwent macular imaging with five OCT devices: HRA Spectralis (Heidelberg Engineering), BMIZAR (TowardPi, 400 kHz swept-source OCTA), TOPCON Maestro 2, NIDEK RS-1 Glauvas, and HUVITZ HOCT-1/1F. Macular thickness was measured in the central 1-mm ETDRS subfield and the 3-mm and 6-mm inner rings (superior, nasal, inferior, and temporal sectors). Three consecutive scans were acquired per device during a single visit by a single experienced operator. Intra-device repeatability was assessed using ICC(3,1) and repeated-measures ANOVA with Bonferroni correction. Inter-device agreement was evaluated using ICC(2,1), Bland-Altman analysis, and absolute percentage error (APE). HRA-anchored comparisons were used for directional bias interpretation, and all 10 direct pairwise device comparisons were additionally performed as cross-validation.
resultsAll devices demonstrated excellent within-device repeatability (ICC > 0.85 across all regions). Inter-device agreement was strongly location-dependent, with PIG: ICC values declining from the central 1-mm subfield (HRA vs. HUVITZ: 0.942; HRA vs. NIDEK: 0.902) to the 6-mm ring (ICC < 0.40 for most pairs). This represents a relative decline of 40-67%. Central thickness ranged from 227.85 μm (TOPCON) to 290.73 μm (BMIZAR). Device-specific systematic biases were identified: BMIZAR overestimated HRA by a mean relative bias of 10.98%, and TOPCON underestimated HRA by a mean relative bias 13.65%, while HUVITZ (+ 2.60%) and NIDEK (+ 3.64%) showed smaller relative deviations (all biases as HRA minus comparator). Limits of agreement widened by a relative 30-50% from central to peripheral rings. Median APE was lowest for HUVITZ (2.39%) and highest for TOPCON (12.23%).
conclusionsMacular thickness measurements are highly repeatable within individual OCT systems but are not interchangeable across devices, with disagreement amplifying progressively toward peripheral macular zones. The Peripheral Instability Gradient demonstrates that cross-device follow-up carries the greatest risk in parafoveal and perifoveal monitoring. To ensure reliable longitudinal assessment, clinicians should maintain device consistency; when switching is unavoidable, device-specific baselines should be established and calibration adjustments applied.