ArticleUltrasonics sonochemistry2026
Ultrasound-assisted recovery of humic substances from municipal digestate using response surface methodology.
Article in Ultrasonics sonochemistry, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Influence of Extraction Conditions on Organic Carbon Recovery from Biogas Digestate.ACS omega · 2026Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
Anaerobic digestion (AD) is widely employed for sewage sludge stabilization, producing substantial volumes of digestate that require further management. This study explores the recovery of high-quality humic substances (HS), specifically humic acids (HA) and fulvic acids (FA), from anaerobically digested sewage sludge (DSS) using ultrasound-assisted extraction (UAE). Implementing this step prior to downstream digestate management positions DSS as a low-cost, sustainable feedstock to produce high-value agronomic biostimulants and soil conditioners. Response surface methodology (RSM) was applied to optimize the UAE parameters, namely ultrasound amplitude (20-100%) and extraction time (5-20 min), to maximise HA yield. A second-order polynomial model demonstrated a strong fit to the experimental data (R
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.