A color Doppler ultrasound-guided protocol for the precise placement of nerve catheters: an anatomical validation study in fresh and fixed cadavers.

Benedikt Büttner, Holger Rosemann, Nilas Zieseniss, Caspar Mewes, José Hinz, Ashham Mansur, Ingo Bergmann

Abstract readValidation Study
In one paragraph

Article in BMC anesthesiology, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Benedikt BüttnerDepartment of Anesthesiology, Emergency and Intensive Care Medicine, University Medical Center Göttingen, Robert-Koch-Str. 40, Göttingen, Lower Saxony, 37099, Germany. benedikt.buettner@med.uni-goettingen.de.
Holger RosemannDepartment of Anesthesiology, Emergency and Intensive Care Medicine, University Medical Center Göttingen, Robert-Koch-Str. 40, Göttingen, Lower Saxony, 37099, Germany.
Nilas ZiesenissDepartment of Anesthesiology, Emergency and Intensive Care Medicine, University Medical Center Göttingen, Robert-Koch-Str. 40, Göttingen, Lower Saxony, 37099, Germany.
Caspar MewesCenter of Anesthesiology and Intensive Care Medicine, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
José HinzDepartment of Anesthesiology, Emergency and Intensive Care Medicine, University Medical Center Göttingen, Robert-Koch-Str. 40, Göttingen, Lower Saxony, 37099, Germany.
Ashham MansurDepartment of Anesthesiology, Emergency and Intensive Care Medicine, University Medical Center Göttingen, Robert-Koch-Str. 40, Göttingen, Lower Saxony, 37099, Germany.
Ingo BergmannDepartment of Anesthesiology, Emergency and Intensive Care Medicine, University Medical Center Göttingen, Robert-Koch-Str. 40, Göttingen, Lower Saxony, 37099, Germany.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundContinuous analgesia via peripheral nerve catheters (PNCs) is highly effective but prone to tip malposition. Direct ultrasound-guided localization of the PNC tip remains challenging, and no universal method is currently established. This cadaver-based study aimed to validate a color Doppler ultrasound-guided protocol for accurate PNC tip placement at the brachial plexus.

methodsEleven ultrasound-guided PNC placements were performed (5 fixed/6 fresh cadavers) according to our specific protocol. PNC tip positioning was repeatedly assessed by visualizing bolus injections via color Doppler ultrasound following 0.5 cm retraction if applicable. Good placement was defined as a color Doppler signal directly adjacent to the brachial plexus concomitant with hydrodissection within its fascial compartment. Verification was obtained through anatomical dissection of the PNC tip in situ and evaluation of the maximal staining site after dye injection via the PNC.

resultsThe diagnostic agreement between the color Doppler signal and anatomical findings was 100%. In fixed cadavers, correct PNC tip placement was confirmed in 100% (5/5) of the specimens. Among the fresh cadavers, 83.3% (5/6) were correctly placed. One secondary malposition was detected by ultrasound and confirmed macroscopically. Qualitative analysis of dye staining revealed good localization in 81.8%, acceptable localization in 9.1%, and poor localization in 9.1% of the samples.

conclusionOur standardized color Doppler ultrasound-guided protocol reliably facilitates precise PNC tip placement in both fixed and fresh cadavers. Using this method, practitioners can accurately verify and adjust PNC tip position in cadavers. This universally applicable approach may translate to clinical practice but warrants further investigation. TRIAL REGISTRATION NUMBER: DRKS00005626 The present validation report was derived from an unpublished study. This study was registered with the German Clinical Trials Register under the clinical trial number DRKS00005626 in July 2015 [Weblink: https://drks.de/search/de/trial/DRKS00005626 ].

Indexed as

Brachial PlexusUltrasonography, Doppler, ColorUltrasonography, InterventionalAgedCadaverFemaleHumansMaleCadaver StudyColor Doppler ultrasoundNerve catheter placementRegional anesthesia

Identifiers

PMID42067791
PMCPMC13135258

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.