Irreversible electroporation ablation with bipolar electrodes: ultrasound findings of ablation zones.
Article in Radiology and oncology, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundThis study aimed to explore the ultrasound and contrast-enhanced ultrasound (CEUS) imaging characteristics and dimensions of ablation lesions after irreversible electroporation of the swine liver and to determine which imaging modality is more suitable for post-procedural follow-up by correlating imaging findings with histopathology. MATERIALS AND
methodsIrreversible electroporation procedures were conducted on three swine with single bipolar electrodes. All procedures were carried out after laparotomy. Twenty-four ablation zones were created under ultrasound guidance. Ultrasound and CEUS evaluations were performed immediately and 48 h after irreversible electroporation. Liver specimens were harvested 48 h after irreversible electroporation for histopathological analysis.
resultsThe ablation area appeared as a hypoechoic, well-defined lesion on ultrasound and showed no enhancement on CEUS immediately after irreversible electroporation. At 48 h, the ablation zone appeared as an inhomogeneous hyperechoic area with a hyperechoic margin and blurred boundaries on ultrasound. CEUS clearly delineated the boundary of the ablation zone and demonstrated centripetal enhancement. The dimensions of the ablation area measured on CEUS 48 h after irreversible electroporation showed the highest correlation with the pathologic ablation zone size (length: r = 0.909, width: r = 0.942, p < 0.001), whereas ultrasound measurements showed the lowest correlation (length: r = 0.676, width: r = 0.842, p < 0.001).
conclusionsCompared with conventional ultrasound, CEUS can accurately measure the dimension of the ablation area, especially 48 h after irreversible electroporation.