ArticleJournal of applied clinical medical physics2026
Clinical implementation and evaluation of a patient-specific surface-guided clearance mapping system for collision avoidance and noncoplanar beam planning.
Article in Journal of applied clinical medical physics, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
11 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
PURPOSE/
objectivesCollision avoidance is critical in external beam radiotherapy to ensure patient safety and plan deliverability. Limited understanding of the collision-free treatment space risks both patient safety and unnecessary exclusion of useful beams-particularly in noncoplanar setups-resulting in suboptimal plans. Conventional methods (manual clearance checks or CT-based assessments, etc.) are either labor-intensive or fail to account for collision-prone anatomy outside the scan. We investigated and clinically implemented a virtual patient-specific clearance mapping system and evaluated its utility as a noncoplanar beam selection tool to improve plan quality. MATERIALS/
methodsThe system integrates full-body, patient-specific surfaces-acquired during simulation using near-infrared imaging-with interactive 3D linac/couch models. Clearance mapping accuracy was validated through phantom measurements and a comparative analysis with manual clearance checks of 60 patients across treatment sites. Workflow efficiency data were reported over three years of clinical implementation. A workflow for patient-specific non coplanar beam selection was proposed and evaluated in 20 lung stereotactic body radiation therapy (SBRT) and 18 breast stereotactic partial breast irradiation (sPBI) cases.
resultsThe clearance mapping accuracy was within ± 1° (gantry/couch rotation) of phantom measurements. For 60 patients, the virtual predictions accurately identified all potential clearance issues, while manual verification missed 5 collision events. Virtual checks saved approximately 15 min of linac and therapist time per plan and eliminated an average 6.2-clinical hour planning delay. With the proposed beam selection workflow, noncoplanar replans for lung SBRT improved target conformality (Paddick Conformity Index from 0.89 to 0.91, p < 0.01) and reduced low dose spillage. For breast sPBI, heart mean dose was lowered (103 cGy to 68 cGy, p < 0.01). Delivery time increased by approx. 30s per plan.
conclusionsThe virtual clearance mapping system outperformed manual verification, streamlined clinical workflow, and could significantly improve plan quality through efficient noncoplanar beam selection. It has replaced manual verification at our institution.