Evidence map›Paper›PMID 41987253›Full record

ArticleJournal of applied clinical medical physics2026

Daily quality assurance, system and procedure-specific commissioning of a bore-based optical surface imaging system for surface-guided radiotherapy (SGRT).

Guang Li, Xiaoning Liu, William Donahue, Shih-Chi Lin, Chengzhu Zhang, Wendy Harris, David Kanchaveli, Seng Boh Lim, Michell Savacool, Michalis Aristophanous and 1 more

Abstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

11 authors.

Guang LiDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, New York City, New York, USA.ORCID https://orcid.org/0000-0002-9022-2883
Xiaoning LiuDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, New York City, New York, USA.
William DonahueDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, New York City, New York, USA.
Shih-Chi LinDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, New York City, New York, USA.ORCID https://orcid.org/0009-0004-5180-070X
Chengzhu ZhangDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, New York City, New York, USA.
Wendy HarrisDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, New York City, New York, USA.
David KanchaveliDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, New York City, New York, USA.ORCID https://orcid.org/0000-0001-7839-4765
Seng Boh LimDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, New York City, New York, USA.ORCID https://orcid.org/0000-0002-4054-8789
Michell SavacoolDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, New York City, New York, USA.
Michalis AristophanousDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, New York City, New York, USA.ORCID https://orcid.org/0000-0001-9117-1515
Laura CervinoDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, New York City, New York, USA.

Funding

X-RAY CRYSTALLOGRAPHYP30CA008748 · NCI · SLOAN-KETTERING INSTITUTE FOR CANCER RES · PI Hongbiao Carl Lekaye · 1985 to 2026
$347.4M
MSK Cancer Center Support Grant NIH P30 CA008748NCI NIH HHS P30 CA008748
6 · The paper itself

Abstract

purposeThis study aimed to commission the AlignRT inBore system in Ethos/Halcyon linear accelerator (Linac) systems for surface-guided radiotherapy (SGRT), including developing a soft collision quality assurance (QA) to pass daily AlignRT QA without daily calibration, ensuring compatibility between SGRT and image-guided radiotherapy (IGRT) setups, establishing a baseline performance in AlignRT motion monitoring, and enabling breast DIBH SGRT using manual beam gating with timing accuracy assessment.

methodsThe AlignRT inBore system contains three ceiling-mounted camera pods for SGRT setup at the simulation isocenter (sim-ISO), and two ring-mounted camera sets on the spring-supported bore cover for motion monitoring at treatment ISO (txt-ISO), after iterative cone-beam computed tomography (iCBCT) setup. The commissioning includes confirming the ring cameras are outside the radiation fields, verifying IEC coordinate consistency between SGRT and IGRT, and ensuring the compatibility of AlignRT daily QA with collision QA. A dot-array plate was used for SGRT calibration and daily QA at both ISO positions. A soft collision check method was developed to minimize the bore-cover displacement by (1) triggering the collision alarm from the rear side, distant from the front ring cameras, and (2) reproducing the rear bore-cover position by marking before the collision check. Clinical staff were trained for the new daily QA procedure, and the average period between calibrations was calculated over the past 12 months. Additionally, the application-specific breast DIBH SGRT procedure was commissioned through an end-to-end test with manual gating, which was retrospectively evaluated for timing accuracy.

resultsThe AlignRT inBore camera ring was moved 2 cm away from the iCBCT field toward the bore front, and the IEC 61217 coordinate system yields consistent SGRT and IGRT setups. AlignRT inBore daily QA can pass using the soft collision procedure without daily calibration, with an average calibration period of 2 weeks, as ad-hoc calibrations are required after open-bore services. The manual gating accuracy is found to be 0.6s ± 0.2s, which can be further reduced if the beam can be proactively turned off before coaching patients to relax from a DIBH.

conclusionThe commissioning of the AlignRT inBore systems was successfully implemented, including developing and implementing the soft collision QA without daily calibration for 2 weeks, updating inBore camera position (2 cm to the bore front), ensuring SGRT-IGRT couch-shift consistency, establishing a baseline performance (< 0.2 mm drift), and allowing SGRT applications on Ethos/Halcyon systems. The application-specific breast DIBH SGRT procedure with manual beam gating is also commissioned with a new retrospective timing analysis for assessing gating accuracy (0.6s ± 0.2s).

Indexed as

Breast NeoplasmsOptical ImagingParticle AcceleratorsPhantoms, ImagingQuality Assurance, Health CareRadiotherapy, Image-GuidedRadiotherapy Planning, Computer-AssistedCalibrationCone-Beam Computed TomographyHumansImage Processing, Computer-AssistedRadiotherapy DosageRadiotherapy, Intensity-Modulatedcalibrationdaily collision check for bore‐based Linacsoptical surface imaging (OSI)QA and commissioningsurface‐guided radiotherapy (SGRT)

Identifiers

PMID41987253
PMCPMC13083584

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.