Commissioning and clinical implementation of Monaco treatment planning system on a compact pencil beam scanning proton therapy system.
Article in Journal of applied clinical medical physics, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
14 authors.
Funding
Abstract
backgroundTreatment planning system (TPS) requires comprehensive commissioning and validation prior to clinical use to ensure accurate dose calculation and safe clinical implementation. PURPOSE: To enable clinical use of the Monaco proton TPS, we commissioned it on a single-room compact proton therapy system and implemented a clinically integrated workflow using daily synthetic CT (sCT) to track fractional dose delivery.
methodsThe Monte Carlo (MC) dose calculation algorithm in Monaco TPS (version 6.1) was commissioned on an IBA Proteus ONE system. Commissioning included data acquisition, beam modeling, and model validation. Data included in-air spot profiles for 70-227 MeV at five planes (+20 cm and +10 cm upstream, isocenter, -10 cm and -20 cm downstream), integral depth-dose curves in water, and absolute monitor unit calibration. Model validation comprised: percent depth-dose in water for monoenergetic beams using a 10 × 10 cm
resultsDistal range (R90, R80, R20) differences between Monaco and measurement were within 1 mm for 70-227.7 MeV. Output differences for all standard cube plans were within 3%. For the IROC brain phantom, the TLD-to-Monaco MC dose ratio was 1.00, with gamma passing rates of 94% (coronal) and 98% (sagittal). sCT-based evaluation showed 99% CTV receiving ≥98% of the prescribed dose across the course.
conclusionsMonaco's proton MC beam model and dose calculation algorithm were commissioned and validated on IBA Proteus ONE for clinical use. The first patient (November, 2024) was successfully treated, with daily dose delivery accuracy monitored via an in-house sCT platform.