ArticleInterdisciplinary cardiovascular and thoracic surgery2026
Evaluation of a Novel Multielectrode Surgical Pulsed Field Ablation Clamp: An Ex-Vivo Vegetal Study.
Article in Interdisciplinary cardiovascular and thoracic surgery, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. An erratum has been issued. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
- Erratum issued
Authors and funding
12 authors.
Funding
Abstract
objectivesAt present, there is no commercially available surgical device to perform cardiac ablation utilizing pulsed field ablation (PFA). We sought to explore the efficacy and requirements of a novel PFA clamp to achieve transmural lesions in a vegetal model of Tuberosum Solanum (TS).
methodsWe tested the PFA clamp at different voltages (1400 V, 2000 V, 2800 V) and numbers of applications (2 or 4) on slabs of TS having different thicknesses (10 mm, 20 mm, 30 mm). Each slab was sliced transversally and the halves put together or kept whole. Every possible combination was tested twice. After 72 hours, we observed the resulting necrotic tissue and calculated its volume. We assessed the determinants of %volume of necrotic tissue, defined as observed volume relative to the ideal volume of necrotic tissue.
resultsA transmural lesion was noted in 68 (94%) of 72 specimens. Using 4 applications, there was a strong linear relationship between voltage and %volume regardless of specimen thickness (R2 = 0.94, R2 = 0.96, R2 = 0.97, all P < .001) but not when using 2 applications (R2 = 0.37, R2 = 0.87, R2 = 0.69 for 10 mm, 20 mm, and 30 mm, respectively, all P < .05). The voltages at which 100% volume was observed using 4 applications-corresponding to the curve intersections-are 958 V, 928 V, and 1044 V for the 10 mm, 20 mm, and 30 mm specimens, respectively.
conclusionsThe epicardial PFA clamp reliably produced transmural lesions in a vegetal model of TS using 2 or 4 applications. With 4 applications, lesion volume demonstrated a predictable, linear relationship with increasing voltage, supporting the device's potential for controlled and consistent ablation.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.