ArticleMagnetic resonance imaging2026
Effect of spatial resolution and heating duration on accuracy of dynamic proton resonance frequency shift temperature measurement with k-space weighted image contrast reconstruction during focused ultrasound.
Article in Magnetic resonance imaging, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors.
Funding
Abstract
purposeSingle-reference variable flip angle Stack-of-Stars (SR-VFA-SoS) thermometry with KWIC reconstruction provides simultaneous aqueous and fatty tissue temperature monitoring. This work evaluates the effect of k-space weighted image contrast (KWIC) reconstruction of a SR-VFA-SoS sequence on the accuracy of both proton resonance frequency shift (PRFS) and T
methodsMagnetic resonance-guided focused ultrasound (MRgFUS) sonications were performed using an asymmetric KWIC-reconstructed single-reference variable flip angle Stack-of-Stars (SR-VFA-SoS) sequence and a validated segmented echo planar imaging (segEPI) thermometry sequence in tissue-mimicking phantoms and an in vivo rabbit model. KWIC windows of projection length 89-377, voxel resolutions of 1.2 mm isotropic and 1.5 × 1.5 × 2.0 mm, three different sets of echo times, and heating durations of 30.2-123.3 s were evaluated. Peak temperature accuracy, calculated cumulative thermal dose (CTD), and temporal lag were assessed.
resultsIn phantoms and the animal model, significant differences in peak temperatures were observed between the SR-VFA-SoS and segEPI thermometry at 1.5 × 1.5 × 2.0 mm voxel size, but not 1.2-mm isotropic. Temperature discrepancies were independent of echo time. Prolonged heating duration with lower power at the larger voxel size resulted in lower peak temperature error. Asymmetric KWIC-reconstructed SR-VFA-SoS peak temperatures were consistently temporally shifted later than segEPI peak temperatures.
conclusionSR-VFA-SoS thermometry with KWIC reconstruction offers promising temporal resolution for simultaneous aqueous and fatty tissue temperature monitoring during MRgFUS. However, temporal blurring due to KWIC reconstruction results in underestimation and temporal shifting of peak temperatures for large voxel sizes and short heating durations. These observations highlight the need for careful KWIC and MRgFUS parameter optimization.