Assessment of LD-V1 radiochromic film low-dose performance for mega-voltage radiotherapy quality assurance.
Article in Journal of applied clinical medical physics, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundHighly modulated radiotherapy delivery demands functional multileaf collimator (MLC) quality assurance (QA) with the ability to measure very low doses during short (<1 s) temporal segments of beam delivery. A radiochromic film with adequate sensitivity in the few-cGy range is therefore needed. LD-V1 film was designed for 2-20 cGy measurements in diagnostic kV energy ranges, but its performance for megavoltage (MV) beam qualities has not been well characterized. PURPOSE: To demonstrate the feasibility of LD-V1 film for low-dose MV measurements and characterize its signal dependence under conditions relevant for clinical QA.
methodsA 6 MV calibration curve was generated using eight dose levels between 0 and 40 cGy and compared with EBT4 film irradiated under identical conditions. Film characteristics evaluated included dose response, scanner and film signal uniformity, scan repeatability, energy dependence, post-exposure signal evolution, scanning-orientation dependence, and one-dimensional (1-D) profile visualization.
resultsLD-V1 demonstrated higher net optical density (OD) and significantly less measurement uncertainty compared to EBT4 at dose levels less than 40 cGy. Significant energy dependence between kV and MV beam qualities was seen. Average net OD differences of 6.3% were seen between landscape and portrait scanning orientations. Post-irradiation signal growth followed a logarithmic trend, stabilizing at ∼24 h. Scan-to-scan variation was <0.1%. Film-signal uniformity was within 3.9%. 1-D profiles showed discernible dose fall-off for doses as low as 10 cGy.
conclusionsLD-V1 demonstrates suitability for low-dose MV measurements and exhibits improved low-dose sensitivity relative to EBT4, indicating its potential value for MLC QA in highly dynamic treatment deliveries. Nonetheless, its pronounced energy and scanning-orientation dependence underscores the need for rigorous and consistent calibration and handling procedures.