Machine Description File-based Monte Carlo Simulation Modeling and its Validation Using Pencil Beam Scanning Proton Therapy Commissioning Data.

Umesh Bharat Gayake, Bhushankumar J Patil, Kantaram Darekar, Gaurav T Bholane, Sanjay D Dhole, Lalit Chaudhary, Siddhartha Laskar

Abstract read
In one paragraph

Article in Journal of medical physics. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Umesh Bharat GayakeDepartment of Physics, MES Abasaheb Garware College, Pune, Maharashtra, India.
Bhushankumar J PatilDepartment of Physics, MES Abasaheb Garware College, Pune, Maharashtra, India.
Kantaram DarekarDepartment of Physics, MES Abasaheb Garware College, Pune, Maharashtra, India.
Gaurav T BholaneDepartment of Physics, Savitribai Phule Pune University, Pune, Maharashtra, India.
Sanjay D DholeDepartment of Physics, Savitribai Phule Pune University, Pune, Maharashtra, India.
Lalit ChaudharyDepartment of Radiation Oncology, Advanced Centre for Treatment Research and Education in Cancer, Homi Bhabha National Institute, Mumbai, Maharashtra, India.
Siddhartha LaskarDepartment of Radiation Oncology, Tata Memorial Hospital, Tata Memorial Centre, Homi Bhabha National Institute, Mumbai, Maharashtra, India.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Purpose: This study aims to detail the machine description file (MDF)-based modeling and validation of the independent Monte Carlo (MC) dose engine using commissioning beam data of a multi-gantry Proteus Plus proton pencil beam scanning (PBS) machine, also, enhancing the understanding of its operational parameters and clinical applications. Materials and Methods: The commissioning of the PBS machine involved measuring key parameters such as integrated depth dose, spot size, and absolute dose using calibrated detectors and ionization chambers. These measurements facilitated the modeling of parameters like energy spread and spot tuning within the RayStation treatment planning system and the tool for particle simulation (TOPAS) Geant4-based MC framework. The validation process employed a MDF containing commissioning beam data, assessing 36 spread-out Bragg peak plans by evaluating range, modulation, distal dose fall-off, and lateral profile characteristics at varying depths. In addition, five patient-specific quality assurance (PSQA) plans were created to evaluate nonhomogeneous fields, following American Association of Physicists in Medicine TG report 185 guidelines. Results: The validation results indicated strong agreement between modeled and measured data, with acceptable comparisons observed between RayStation and TOPAS outputs for homogeneous fields. Nonhomogeneous PSQA results demonstrated that over 95% of cases met the gamma passing criteria of 3%/3 mm. Furthermore, assessments of water equivalent thickness in heterogeneous tissues - such as muscle, fat, bone, and liver - showed discrepancies within 0.5% of the range uncertainty, confirming the accuracy of the modeling approach. Conclusion: The MDF-based modeling and validation process is a robust and practical approach for clinical environments, effectively supporting proton therapy research and ensuring quality assurance in treatment planning. The methodologies developed in this study provide valuable insights and tools for researchers utilizing TOPAS MC as a clinical research instrument, paving the way for advancements in proton therapy techniques and patient-specific treatment plans.

Indexed as

Monte Carlopencil beamproton therapysimulationtool for particle simulationtreatment planning system