Evidence map›Paper›PMID 41923382›Full record

ArticleProceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine2026

Measurement of the wear volumes generated within retrieved magnetically controlled growing rods.

Thomas J Joyce

Abstract read
In one paragraph

Article in Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

1 author.

Thomas J JoyceDepartment of Mechanical and Construction Engineering, Northumbria University, Newcastle, UK.ORCID 0000-0002-8497-5790

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Magnetically controlled growing rods are used for the surgical treatment of early onset scoliosis. However, one of the key concerns around the use of these rods is the extensive titanium alloy metallosis seen at revision surgery. To quantify the wear causing this metallosis, for the first time the internal wear volumes from magnetically controlled growing rods have been measured. A cohort of 30 retrieved rods was measured, from 18 cases (6 single rods and 12 dual rod constructs) after an average time in vivo of 1120 days. Another pair of rods explanted after 30 days due to infection served as a negligible-wear datum to validate the methodology. From each retrieved rod, the wear volume of the extending bar component and of the outer casing component was measured. This was achieved through geometric measurements of the components, with the geometry of the internal aspect of the outer casings determined using a replicating compound and then measuring the replica. The average wear volume of the retrieved rods was 71.8 (21.2-201.9) mm

Indexed as

Magnetic PhenomenaMagneticsMaterials TestingMechanical PhenomenaAlloysHumansTitaniumAlloysTitaniumexplant analysismagnetically controlled growing rodretrieval studyscoliosistitanium alloywear

Identifiers

PMID41923382
PMCPMC13199638

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.