ArticleJournal of cataract and refractive surgery2026
Occlusion break surge performance of a next-generation phacoemulsification system.
Article in Journal of cataract and refractive surgery, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeTo compare occlusion break surge volumes of UNITY Vitreoretinal Cataract System (UVCS) with CENTURION Vision System with Active Sentry (CAS).
settingAlcon Research, LLC., Lake Forest, California.
designExperimental study.
methodsAn eye model representing average human eye compliance was used to determine average occlusion break surge volumes. UVCS was tested at target intraocular pressures (IOPs) of 20 mm Hg, 30 mm Hg, 40 mm Hg, 50 mm Hg, 60 mm Hg, and 80 mm Hg; vacuum limits of 300 mm Hg, 400 mm Hg, 500 mm Hg, 600 mm Hg, and 650 mm Hg; and aspiration flow rates of 20 cc/min and 40 cc/min. CAS was tested at target IOPs of 30 mm Hg, 40 mm Hg, 50 mm Hg, 60 mm Hg, and 80 mm Hg, and at the same set vacuum limits and aspiration flow rates as UVCS. In addition, a second-order polynomial was fit to the entire dataset to compare UVCS and CAS at a target IOP of 20 mm Hg.
resultsAfter calculating sample means at each setpoint combination, the largest mean surge volume was 48 μL for UVCS and 75 μL for CAS, corresponding to 19% and 30%, respectively, of the average human phakic anterior chamber volume. For all common setpoint combinations, mean occlusion break surge volumes were 44% lower for UVCS than CAS. At a flow rate of 40 cc/min and IOP setpoint of 20 mm Hg, UVCS showed a 46% average decrease across all vacuum limits.
conclusionsUVCS demonstrated lower mean surge volumes after occlusion break than CAS. This may give surgeons greater confidence operating at lower IOPs and may reduce the risk of posterior capsule rupture, especially in complex cases.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials