ArticleJournal of perinatology : official journal of the California Perinatal Association2026
Comparison of nasal interfaces and pressures during critical time-points in NICU: observation data from over 10 years.
Article in Journal of perinatology : official journal of the California Perinatal Association, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectiveThe pressure amount and fluctuations delivered to the nasopharynx by bubble CPAP (BCPAP) is affected by the nasal interface, but the clinical effects of transmitted pressure fluctuations from BCPAP are unknown. STUDY
designRetrospective study of 2432 infants < 33 weeks GA had respiratory variables electronically extracted for: 1) delivery room, 2) initial 72 h, 3) after extubation, and 4) 31 weeks PMA. BCPAP nasal interfaces (RAM® cannula and Babi.Plus
resultsRAM5/6 had more CPAP failures than RAM7/8 or Babi5/6. Babi5/6 had less CPAP failure in first 72 h in NICU, but RAM7/8 had less in DR. Less infants failed CPAP at extubation in Babi5/6 due to decreased apnea events.
conclusionAt similar transmitted nasopharyngeal pressures, Babi.Plus had less CPAP failure and apnea than RAM in NICU, whereas the ease of application of RAM may help in delivery room.
Identifiers
41912704What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.