Evidence map›Paper›PMID 41893170›Full record

ArticleJournal of functional biomaterials2026

The Effect of Biomechanical Loading Parameters on the Stress and Strain Behavior of Orthodontic Mini-Implants: A Finite Element Study.

Tinela Panaite, Cristian Liviu Romanec, Bogdan Radu Dragomir, Ana Sîrghie, Carmen Amititeloaie, Carina Balcos, Carmen Diana Nicoleta Savin

Abstract read
In one paragraph

Article in Journal of functional biomaterials, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Tinela PanaiteGrigore T. Popa University of Medicine and Pharmacy Iasi, 700259 Iași, Romania.ORCID 0009-0002-8042-2424
Cristian Liviu RomanecGrigore T. Popa University of Medicine and Pharmacy Iasi, 700259 Iași, Romania.
Bogdan Radu DragomirGrigore T. Popa University of Medicine and Pharmacy Iasi, 700259 Iași, Romania.
Ana SîrghieGrigore T. Popa University of Medicine and Pharmacy Iasi, 700259 Iași, Romania.
Carmen AmititeloaieGrigore T. Popa University of Medicine and Pharmacy Iasi, 700259 Iași, Romania.ORCID 0009-0003-8065-015X
Carina BalcosGrigore T. Popa University of Medicine and Pharmacy Iasi, 700259 Iași, Romania.ORCID 0000-0001-7712-1319
Carmen Diana Nicoleta SavinGrigore T. Popa University of Medicine and Pharmacy Iasi, 700259 Iași, Romania.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

BACKGROUND/

objectivesThis study evaluated the influence of key biomechanical parameters-orthodontic force magnitude, loading direction, and insertion depth-on stress and strain distribution in orthodontic mini-implants using three-dimensional finite element analysis (FEM).

methodsA three-dimensional model of a titanium orthodontic mini-implant inserted into a mandibular bone segment was developed and analyzed under varying force magnitudes (1-10 N), loading directions (30°, 45°, and 60°), and insertion depths (2-4 mm). Cortical and cancellous bone components were included, and static loading conditions were applied using simplified, linear elastic material assumptions.

resultsStress and strain levels increased with higher force magnitudes, with implant stresses approaching critical values at loads above 9 N. Cortical bone stresses remained within physiological limits, whereas cancellous bone exceeded the microdamage strain threshold at forces greater than 3 N. A 60° loading direction reduced implant bending and strain, while deeper insertion significantly decreased strain and displacement, indicating improved primary stability.

conclusionsWithin the limits of this computational model, optimal mechanical behavior was observed under 1-3 N forces, a 60° loading direction, and a 2-4 mm insertion depth. Loads above 9 N approached fatigue and interfacial risk. These findings provide computational insight into the biomechanical behavior of orthodontic mini-implants under the modeled conditions.

Indexed as

anchorage stabilityfinite element analysisinsertion depthloading directionorthodontic forceorthodontic mini-implantsprimary stability

Identifiers

PMID41893170
PMCPMC13027284

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.