Implant-friendly MRI of deep brain stimulation patients at 3T.
Article in NeuroImage, 2026. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
Abstract
introductionMRI of patients with deep brain stimulation (DBS) implants presents a significant safety risk due to RF heating. A current measurement that utilizes an inherent imaging artifact was previously used to mitigate induced current along a single electrode using a two transmit-channel system. This study extends it to bilateral DBS configurations and demonstrates a streamlined implant-friendly (IF) optimization in both unilateral and bilateral patients.
methodsThe bilateral IF mode was optimized to reduce RF-induced currents on both electrodes while maximizing the overall B1 and was validated in an anthropomorphic head phantom. An imaging workflow to determine electrode-specific and bilateral IF modes were demonstrated in three DBS patients (one bilateral, two unilateral). Image quality and artifact reduction of the circularly-polarized (CP) and IF modes were compared using 2D-TSE.
resultsPhantom study confirmed the heating reduction capabilities of the IF optimization. In both phantom and patient studies, RF artifact reduction was observed. In the unilateral DBS case, IF optimization completely mitigated the induced current, while in the bilateral case, current was reduced up to 74 % compared to CP excitation. The total active scan time for the pre-scan imaging was <6 min. All measurements in the pre-scan protocol were performed directly on the MRI console, with minimal external hardware.
conclusionThis study demonstrates a practical workflow for real-time current analysis and IF mode optimization in patients with DBS implants, effectively reducing RF-induced heating and artifacts while maintaining imaging performance.